[单枚拉力螺钉前路固定术后齿突弯曲刚度:生物力学研究]

[Odontoid bending stiffness after anterior fixation with a single lag screw: biomechanical study].

作者信息

Buchvald P, Čapek L, Barsa P

机构信息

Neurocentrum, Krajská nemocnice Liberec, a. s.

出版信息

Acta Chir Orthop Traumatol Cech. 2015;82(3):235-8.

Abstract

PURPOSE OF THE STUDY The aim of the experiment was to compare the bending stiffness of an intact odontoid process with bending stiffness after its simulated type II fracture was fixed with a single lag screw. The experiment was done with a desire to answer the question of whether a single osteosynthetic screw is sufficient for good fixation of a type II odontoid fracture. MATERIAL AND METHODS The C2 vertebrae of six cadavers were used. With simultaneous measurement of odontoid bending stiffness, the occurrence of a fracture (type IIA, Grauer's modification of the Anderson- D'Alonzo classification) was simulated using action exerted by a tearing machine in the direction perpendicular to the odontoid axis. Each odontoid fracture was subsequently treated by direct osteosynthesis with a single lag screw inserted in the axial direction by a standard surgical procedure in order to provide conditions similar to those achieved by routine surgical management. The treated odontoid process was subsequently subjected to the same tearing machine loading as applied to it at the start of the experiment. The bending stiffness measured was then compared with that found before the fracture occurred. The results were statistically evaluated by the t-test for paired samples at the level of significance α = 0.05. RESULTS The average value of bending stiffness for odontoid processes of intact vertebrae at the moment of fracture occurrence was 318.3 N/mm. After single axial lag screw fixation of the fracture, the average bending stiffness for the odontoid processes treated was 331.3 N/mm. DISCUSSION Higher values of bending stiffness after screw fixation were found in all specimens and, in comparison with the values recorded before simulated fractures, the increase was statistically significant. CONCLUSIONS The results of our measurements suggest that the single lag screw fixation of a type IIA odontoid fracture will provide better stability for the fracture fragment-C2 body complex on antero-posterior perpendicular loading than can be found in intact C2 vertebrae. Key words: odontoid fracture, odontoid fixation, bending stiffness, lag screw.

摘要

研究目的 本实验旨在比较完整齿突的弯曲刚度与模拟II型骨折用单枚拉力螺钉固定后的弯曲刚度。进行该实验是为了回答单枚骨合成螺钉是否足以良好固定II型齿突骨折这一问题。材料与方法 使用六具尸体的C2椎体。在同时测量齿突弯曲刚度的情况下,使用撕裂机在垂直于齿突轴线的方向施加作用力来模拟骨折(IIA 型,Grauer对Anderson-D’Alonzo分类法的改良)。随后,通过标准手术操作沿轴向插入单枚拉力螺钉进行直接骨合成来治疗每例齿突骨折,以提供与常规手术治疗相似的条件。随后,对治疗后的齿突施加与实验开始时相同的撕裂机负荷。然后将测得的弯曲刚度与骨折发生前的弯曲刚度进行比较。结果采用配对样本t检验在显著性水平α = 0.05下进行统计学评估。结果 骨折发生瞬间完整椎体齿突的弯曲刚度平均值为318.3 N/mm。骨折用单枚轴向拉力螺钉固定后,治疗后的齿突平均弯曲刚度为331.3 N/mm。讨论 在所有标本中均发现螺钉固定后弯曲刚度值更高,与模拟骨折前记录的值相比,增加具有统计学意义。结论 我们的测量结果表明,IIA型齿突骨折的单枚拉力螺钉固定在前后垂直负荷下为骨折块-C2椎体复合体提供的稳定性优于完整C2椎体。关键词:齿突骨折;齿突固定;弯曲刚度;拉力螺钉

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验