Suppr超能文献

两枚Gamma 3髓内钉失败后行挽救性全髋关节置换术:1例病例报告

Salvage total hip arthroplasty after two gamma 3 nail failures: a case report.

作者信息

Stepanovic Z L, Ristic B M

机构信息

Clinic for Orthopaedic and Trauma Surgery, Clinical Center Kragujevac, Serbia.

出版信息

Acta Chir Orthop Traumatol Cech. 2015;82(3):239-42.

Abstract

We report a case of 68-year-old female patient who suffered two Gamma 3 nail breakages following trochanteric fixation failure revision surgery. A patient was primarily treated with sliding hip screw. Revision surgeries of reverse trochanteric nonunion were performed with short Gamma 3 nails and bone grafting. The nail failures occurred 11 and 7 months after implantation. This undesired series of events were caused by inappropriate implant selection, thus replacing the nail to a modular total hip arthroplasty was more effective as salvage surgery. Key words: salvage total hip arthroplasty, gamma nail, sliding hip screw, fixation failure, reverse trochanteric fracture, nonunion.

摘要

我们报告一例68岁女性患者,其在转子间固定失败翻修手术后发生了两次Gamma 3钉断裂。该患者最初接受动力髋螺钉治疗。采用短Gamma 3钉和植骨进行了转子间骨不连的翻修手术。钉子在植入后11个月和7个月发生断裂。这一系列不良事件是由不恰当的植入物选择导致的,因此,作为挽救手术,将钉子更换为模块化全髋关节置换术更为有效。关键词:挽救性全髋关节置换术、Gamma钉、动力髋螺钉、固定失败、转子间骨折、骨不连