Amirouche Farid, Solitro Giovanni F, Chandrasekaran Siva, Domb Benjamin G, Gonzalez Mark H
Department of Orthopaedics, University of Illinois at Chicago, Chicago, Illinois.
Advocate Health Care, Downers Grove, Illinois.
J Arthroplasty. 2016 Jan;31(1):323-9. doi: 10.1016/j.arth.2015.07.042. Epub 2015 Aug 4.
This study aims to validate a modified circle theorem method for the calculation of true version of the acetabular component on anteroposterior x-rays with intra-operative version data derived from robotic assisted total hip arthroplasty (THA). Planar anteversion measurements recorded intraoperatively in 80 THAs were correlated to measurements on anteroposterior radiographs. The mean anteversion of the cohort measured by the robotic system and on plain radiography was 21.2° ± 2.0° and 19.9° ± 3.4° respectively and 97.5% of cases were in a 30% relative error. The correlation between the true and planar measurements of anteversion on plain radiographs was strong (Pearson correlation coefficient of 0.9422). We conclude that the circle theorem method can be validated with data from robotic guided THA.
本研究旨在验证一种改良的圆定理方法,该方法用于根据机器人辅助全髋关节置换术(THA)术中获得的髋臼组件前倾角数据,计算前后位X线片上髋臼组件的真实前倾角。对80例全髋关节置换术中记录的术中平面前倾角测量值与前后位X线片测量值进行相关性分析。该队列中通过机器人系统测量的平均前倾角和X线平片测量的平均前倾角分别为21.2°±2.0°和19.9°±3.4°,97.5%的病例相对误差在30%以内。X线平片上前倾角的真实测量值与平面测量值之间的相关性很强(Pearson相关系数为0.9422)。我们得出结论,圆定理方法可以通过机器人引导的全髋关节置换术数据得到验证。