使用具有先进导管定位功能的接触力感应消融导管可显著减少房颤导管消融术中的透视时间和辐射剂量。

Use of a contact force-sensing ablation catheter with advanced catheter location significantly reduces fluoroscopy time and radiation dose in catheter ablation of atrial fibrillation.

作者信息

Lee Geoffrey, Hunter Ross J, Lovell Matthew J, Finlay Malcom, Ullah Waqas, Baker Victoria, Dhinoja Mehul B, Sporton Simon, Earley Mark J, Schilling Richard J

机构信息

Cardiovascular Biomedical Research Unit, St Bartholomew's Hospital, Barts Health NHS Trust, First Floor Dominion House, 60 Bartholomew Close, West Smithfield, London EC1A 7BE, UK Department of Cardiology, The Royal Melbourne Hospital and The Department of Medicine, University of Melbourne, Australia.

Cardiovascular Biomedical Research Unit, St Bartholomew's Hospital, Barts Health NHS Trust, First Floor Dominion House, 60 Bartholomew Close, West Smithfield, London EC1A 7BE, UK.

出版信息

Europace. 2016 Feb;18(2):211-8. doi: 10.1093/europace/euv186. Epub 2015 Aug 27.

Abstract

AIMS

The aim of this study was to evaluate the 'real-world' impact of a novel contact force (CF)-sensing (SmartTouch™, Biosense Webster, Diamond Bar, CA, USA) catheter coupled with an advanced catheter location (ACL) system on fluoroscopy time and fluoroscopy dose during atrial fibrillation (AF) ablation.

METHODS AND RESULTS

This was a retrospective observational cohort study of prospectively collected data of 1515 consecutive patients undergoing paroxysmal AF (PAF) and persistent AF (PerAF) ablation at a single institution between 2009 and 2014. Patients undergoing AF ablation with the SmartTouch catheter and the ACL system (SmartTouch group, n = 510) were compared with those undergoing AF ablation without this technology (control group, n = 1005). The primary outcomes were total fluoroscopy time (min) and fluoroscopy dose as measured by the dose-area product (mGy cm(2)). Secondary endpoints included total procedure time, total ablation time, and major cardiac complications (tamponade, pericardial effusion, and urgent cardiac surgery). The SmartTouch group had significantly lower fluoroscopy times (9.5 vs. 41 min, P < 0.001), radiation doses (1044 vs. 3571 mGy cm(2), P < 0.001), and shorter procedural time (195 vs. 240 min, P < 0.001) when compared with the control group. This was statistically significant for both PAF and PerAF ablations and for both de novo and redo AF procedures. After a learning curve, a median fluoroscopy time of 3.5 min (interquartile range 6) for all AF ablations was achieved. There was no difference in the rate of cardiac complications (∼ 1.5%).

CONCLUSION

SmartTouch™ CF-sensing catheter use with ACL™ during AF ablation significantly reduces fluoroscopy times by 77%, radiation dose by 71%, and procedural time by 19% but does not improve overall safety or the risk of cardiac complications.

摘要

目的

本研究旨在评估一种新型接触力(CF)传感导管(SmartTouch™,美国加利福尼亚州钻石吧市百盛韦伯斯特公司)与先进导管定位(ACL)系统联用,对心房颤动(AF)消融术中透视时间和透视剂量的“真实世界”影响。

方法与结果

这是一项回顾性观察队列研究,对2009年至2014年间在单一机构接受阵发性AF(PAF)和持续性AF(PerAF)消融的1515例连续患者的前瞻性收集数据进行分析。将使用SmartTouch导管和ACL系统进行AF消融的患者(SmartTouch组,n = 510)与未使用该技术进行AF消融的患者(对照组,n = 1005)进行比较。主要结局指标为总透视时间(分钟)和用剂量面积乘积(mGy cm²)测量的透视剂量。次要终点包括总手术时间、总消融时间和主要心脏并发症(心包填塞、心包积液和紧急心脏手术)。与对照组相比,SmartTouch组的透视时间显著缩短(9.5分钟对41分钟,P < 0.001),辐射剂量显著降低(1044 mGy cm²对3571 mGy cm²,P < 0.001),手术时间显著缩短(195分钟对240分钟,P < 0.001)。这在PAF和PerAF消融以及初发和再次AF手术中均具有统计学意义。经过学习曲线后,所有AF消融的中位透视时间达到3.5分钟(四分位间距6)。心脏并发症发生率无差异(约1.5%)。

结论

在AF消融术中使用SmartTouch™ CF传感导管与ACL™联用可使透视时间显著缩短77%,辐射剂量显著降低71%,手术时间显著缩短19%,但并未提高总体安全性或降低心脏并发症风险。