注塑成型氧化锆种植体去除扭矩分析;对比格犬的实验研究。

Analysis of Removal Torque of Injection Molded Zirconia Implants; An Experimental Study on Beagles.

作者信息

Oh Gye-Jeong, Ban Jae-Sam, Lim Hyun-Pil, Yun Kwi-Dug, Lee Kwang-Min, Vang Mong-Sook, Yang Hong-So, Kang Seong-Soo, Shin Jin-Ho, Kim Ga-Hyun, Ji Min-Kyung, Park Sang-Won, Fisher John G

出版信息

J Nanosci Nanotechnol. 2015 Jan;15(1):339-41. doi: 10.1166/jnn.2015.8376.

Abstract

This study compared the removal torque between injection molded zirconia implants and titanium implants with resorbable blast media (RBM) surfaces in beagle humeri. Fifteen screw-shaped implants were classified into 3 groups; titanium implant with RBM surface (Group RT), injection molded zirconia implant (Group Zr) and injection molded zirconia implant with sand-blasted surface (Group ZrS). Implants were inserted into beagle humeri. After 12 weeks, removal torque values were measured. The Zr group has a slightly higher removal torque value than the RT and ZrS groups but there were no significant differences among groups. Zirconia implants shows a similar removal torque to RBM titanium implants. This in vivo study showed injection molded zirconia implants could be an alternative to RBM titanium implants in terms of removal torque.

摘要

本研究比较了注射成型氧化锆种植体与具有可吸收喷砂介质(RBM)表面的钛种植体在比格犬肱骨中的拆除扭矩。将15枚螺旋形种植体分为3组;具有RBM表面的钛种植体(RT组)、注射成型氧化锆种植体(Zr组)和具有喷砂表面的注射成型氧化锆种植体(ZrS组)。将种植体植入比格犬肱骨。12周后,测量拆除扭矩值。Zr组的拆除扭矩值略高于RT组和ZrS组,但各组之间无显著差异。氧化锆种植体的拆除扭矩与RBM钛种植体相似。这项体内研究表明,就拆除扭矩而言,注射成型氧化锆种植体可以替代RBM钛种植体。