Lundström H
Department of Building, Energy and Environmental Engineering, University of Gävle, SE-801 76 Gävle, Sweden.
Rev Sci Instrum. 2015 Aug;86(8):086104. doi: 10.1063/1.4929375.
Annealing procedures for hot-wire sensors of platinum and platinum-plated tungsten have been investigated experimentally. It was discovered that the two investigated sensor metals behave quite differently during the annealing process, but for both types annealing may improve long-term stability considerably. Measured drift of sensors both without and with prior annealing is presented. Suggestions for suitable annealing temperatures and times are given.
对铂和镀铂钨热线传感器的退火程序进行了实验研究。结果发现,在退火过程中,所研究的两种传感器金属表现出截然不同的特性,但对于这两种类型的传感器,退火均可显著提高其长期稳定性。文中给出了未经退火和经过预先退火的传感器的测量漂移情况。还给出了合适的退火温度和时间的建议。