Kim Kug Jin, Song Eun Kyoo, Seon Jong Keun, Seol Jong Hwan
Department of Orthopedic Surgery, Wolchul Ki-Chan Hospital, Yeongam, Korea.
Department of Orthopedic Surgery, Center for Joint Disease, Chonnam National University Hwasun Hospital, Hwasun, Korea.
Knee Surg Relat Res. 2015 Sep;27(3):181-6. doi: 10.5792/ksrr.2015.27.3.181. Epub 2015 Sep 1.
The purpose of this study was to compare the mechanical stability of three types of plate systems for opening wedge high tibial osteotomy.
Forty-eight fresh frozen porcine tibia specimens were assigned to three different fixation device groups: Aescular group (16 specimens) was fixed with Aescular plates; Puddu group (16 specimens) with a Puddu plate, and TomoFix group (16 specimens) with a TomoFix plate. We compared axial displacements under compression loads from 200 to 2,000 N and maximal loads at failure among 8 specimens per group. We also compared displacements under cyclic load after 100 cycles at a compressive load of 2,000 N among 8 specimens per group.
In all three groups, displacement under compression load increased with the increase in the axial compressive load; however, no significant intergroup differences were observed in the mean values under tested loading conditions. The mean maximal loads at failure were not significantly different (6,055, 6,798, and 6,973 N in the Aescular, Puddu, and TomoFix groups, respectively; p=0.41). While the TomoFix group showed less extension and strain during the cyclic load test, the mean values showed no significant differences among groups.
All three plate systems were found to provide fixation stability suitable for bearing axial compression and cyclic loads while walking.
本研究旨在比较三种用于开放性楔形高位胫骨截骨术的钢板系统的机械稳定性。
48个新鲜冷冻猪胫骨标本被分配到三个不同的固定装置组:Aescular组(16个标本)用Aescular钢板固定;Puddu组(16个标本)用Puddu钢板固定,TomoFix组(16个标本)用TomoFix钢板固定。我们比较了每组8个标本在200至2000 N压缩载荷下的轴向位移以及失效时的最大载荷。我们还比较了每组8个标本在2000 N压缩载荷下进行100次循环后的循环载荷下的位移。
在所有三组中,压缩载荷下的位移随轴向压缩载荷的增加而增加;然而,在测试载荷条件下,各组平均值之间未观察到显著的组间差异。平均最大失效载荷无显著差异(Aescular组、Puddu组和TomoFix组分别为6055、6798和6973 N;p = 0.41)。虽然TomoFix组在循环载荷试验中显示出较小的伸展和应变,但各组平均值之间无显著差异。
发现所有三种钢板系统都能提供适合行走时承受轴向压缩和循环载荷的固定稳定性。