Silverstein Evan, Donahue Sean P
J Pediatr Ophthalmol Strabismus. 2015 Nov-Dec;52(6):364-70. doi: 10.3928/01913913-20150915-01. Epub 2015 Sep 22.
Automated vision screeners can identify children with amblyopia risk factors. Two screening instruments having different referral criteria were evaluated in a community setting: SPOT (Pediavision, Lake Mary, FL) (sensitive manufacturer's referral criteria) and plusoptiX S08 (Plusoptix GmbH, Nuremberg, Germany) (specific modified Arthur referral criteria).
All children were screened by SPOT, and referred children were then screened using plusoptiX. Referred children received a gold standard examination to determine whether amblyopia risk factors were present.
A total of 2,801 treatment-naïve children were screened using SPOT. Of these, 307 (11.0%) were referred by SPOT and subsequently screened by plusoptiX; 100 received a gold standard examination. Amblyopia risk factors were present in 43% (43 of 100) referred by SPOT compared to 72.7% (32 of 44) for plusoptiX. Eleven of 56 referred by SPOT had amblyopia risk factors that would have been missed by plusoptiX, including three with mild amblyopia.
PlusoptiX with modified Arthur referral criteria can be a highly specific screening device detecting amblyopia risk factors without missing children with moderate/severe amblyopia.
自动视力筛查仪可识别有弱视风险因素的儿童。在社区环境中对两种具有不同转诊标准的筛查仪器进行了评估:SPOT(佛罗里达州玛丽湖的Pediavision公司)(制造商敏感的转诊标准)和plusoptiX S08(德国纽伦堡的Plusoptix GmbH公司)(特定修改后的亚瑟转诊标准)。
所有儿童均通过SPOT进行筛查,然后对转诊儿童使用plusoptiX进行筛查。转诊儿童接受金标准检查以确定是否存在弱视风险因素。
共有2801名未接受过治疗的儿童通过SPOT进行了筛查。其中,307名(11.0%)被SPOT转诊,随后通过plusoptiX进行筛查;100名接受了金标准检查。SPOT转诊的儿童中有43%(100名中的43名)存在弱视风险因素,而plusoptiX为72.7%(44名中的32名)。SPOT转诊的56名儿童中有11名存在plusoptiX会漏诊的弱视风险因素,包括3名轻度弱视儿童。
采用修改后亚瑟转诊标准的plusoptiX可以成为一种高度特异的筛查设备,能检测出弱视风险因素,且不会漏诊中度/重度弱视儿童。