In Vivo Wear Performance of Cobalt-Chromium Versus Oxidized Zirconium Femoral Total Knee Replacements.

作者信息

Gascoyne Trevor C, Teeter Matthew G, Guenther Leah E, Burnell Colin D, Bohm Eric R, Naudie Douglas R

机构信息

Orthopaedic Innovation Centre, Winnipeg, MB, Canada.

London Health Sciences Centre, London, ON, Canada.

出版信息

J Arthroplasty. 2016 Jan;31(1):137-41. doi: 10.1016/j.arth.2015.07.045. Epub 2015 Aug 29.

Abstract

This study examines the damage and wear on the polyethylene (PE) inserts from 52 retrieved Genesis II total knee replacements to identify differences in tribological performance between matched pairs of cobalt-chromium (CoCr) and oxidized zirconium (OxZr) femoral components. Observer damage scoring and microcomputed tomography were used to quantify PE damage and wear, respectively. No significant differences were found between CoCr and OxZr groups in terms of PE insert damage, surface penetration, or wear. No severe damage such as cracking or delamination was noted on any of the 52 PE inserts. Observer damage scoring did not correlate with penetrative or volumetric PE wear. The more costly OxZr femoral component does not demonstrate clear tribological benefit over the standard CoCr component in the short term with this total knee replacement design.

摘要

本研究检查了52例回收的Genesis II全膝关节置换术中聚乙烯(PE)衬垫的损伤和磨损情况,以确定配对的钴铬(CoCr)和氧化锆(OxZr)股骨部件之间的摩擦学性能差异。分别采用观察者损伤评分和微型计算机断层扫描来量化PE损伤和磨损。在PE衬垫损伤、表面穿透或磨损方面,CoCr组和OxZr组之间未发现显著差异。52个PE衬垫中均未发现诸如开裂或分层等严重损伤。观察者损伤评分与PE穿透性或体积磨损无关。对于这种全膝关节置换设计,在短期内,成本更高的OxZr股骨部件相对于标准CoCr部件并未显示出明显的摩擦学优势。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验