Suppr超能文献

两种商用甲氨蝶呤免疫分析方法分析性能的直接比较以及与液相色谱-质谱联用和原荧光偏振免疫分析的对比

Head to head evaluation of the analytical performance of two commercial methotrexate immunoassays and comparison with liquid chromatography-mass spectrometry and the former fluorescence polarization immunoassay.

作者信息

Günther Viola, Mueller Daniel, von Eckardstein Arnold, Saleh Lanja

出版信息

Clin Chem Lab Med. 2016 May;54(5):823-31. doi: 10.1515/cclm-2015-0578.

Abstract

BACKGROUND

Monitoring of plasma drug levels is mandatory in patients receiving high-dose methotrexate. This study evaluated the analytical performance of the novel Architect and the established ARK™ methotrexate immunoassay (running on the Roche Cobas© c502 analyzer) in comparison with liquid chromatography-mass spectrometry (LC-MS) and the TDx/TDxFLx Methotrexate II assay.

METHODS

Imprecision and linearity were verified for the Architect and ARK assay according to CLSI EP15-A3 and EP6-A guidelines, respectively. The reported limit of quantitation (0.04 μmol/L) was tested for both assays according to the CLSI EP17-A2 guideline. Correlation and agreement between the different assays were evaluated using residual plasma samples (n=153).

RESULTS

Total imprecision was <6.3% and <9.5% for the Architect and ARK assay, respectively. The claimed linearity and limit of quantitation were confirmed for the Architect assay. For the ARK assay, imprecision at the limit of quantitation was <18% with a positive bias resulting in a high total error up to 58%, and hence the linearity could not be confirmed. Both assays showed strong correlations with the TDX assay and LC-MS but a positive bias of 12.2% and 20.5% in comparison to LC-MS for the Architect and ARK assay, respectively. For the ARK assay this bias increased dramatically for samples with concentrations towards the limit of quantitation.

CONCLUSIONS

The Architect assay is suitable for monitoring plasma methotrexate, but the ARK assay showed unsatisfactory performance in the analysis of low concentrated samples. Unlike the TDX assay, both assays require manual dilution of samples at higher concentrations, which delays sample processing in clinical routine.

摘要

背景

接受大剂量甲氨蝶呤治疗的患者必须监测血浆药物水平。本研究评估了新型Architect和已有的ARK™甲氨蝶呤免疫分析方法(在罗氏Cobas© c502分析仪上运行)与液相色谱 - 质谱联用(LC-MS)以及TDx/TDxFLx甲氨蝶呤II分析方法相比的分析性能。

方法

分别根据CLSI EP15-A3和EP6-A指南验证了Architect和ARK分析方法的不精密度和线性。根据CLSI EP17-A2指南对两种分析方法报告的定量限(0.04 μmol/L)进行了测试。使用残留血浆样本(n = 153)评估了不同分析方法之间的相关性和一致性。

结果

Architect和ARK分析方法的总不精密度分别<6.3%和<9.5%。Architect分析方法的线性和定量限得到了确认。对于ARK分析方法,定量限处的不精密度<18%,但存在正偏差,导致总误差高达58%,因此无法确认其线性。两种分析方法与TDX分析方法和LC-MS均显示出强相关性,但与LC-MS相比,Architect和ARK分析方法分别存在12.2%和20.5%的正偏差。对于ARK分析方法,浓度接近定量限的样本偏差显著增加。

结论

Architect分析方法适用于监测血浆甲氨蝶呤,但ARK分析方法在低浓度样本分析中表现不佳。与TDX分析方法不同,这两种分析方法在较高浓度时都需要手动稀释样本,这在临床常规中会延迟样本处理。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验