Strudley Cecilia J, Young Kenneth C, Looney Padraig, Gilbert Fiona J
1 National Co-ordinating Centre for the Physics of Mammography, Department of Medical Physics, Royal Surrey County Hospital NHS Foundation Trust, Guildford, UK.
2 Department of Physics, University of Surrey, Guildford, UK.
Br J Radiol. 2015;88(1056):20150324. doi: 10.1259/bjr.20150324. Epub 2015 Oct 14.
To develop tomosynthesis quality control (QC) test methods and use them alongside established two-dimensional (2D) QC tests to measure the performance of digital breast tomosynthesis (DBT) systems used in a comparative trial with 2D mammography.
DBT QC protocols and associated analysis were developed, incorporating adaptions of some 2D tests as well as some novel tests. The tomosynthesis tests were: mean glandular dose to the standard breast model; contrast-to-noise ratio in reconstructed focal planes; geometric distortion; artefact spread; threshold contrast detail detection in reconstructed focal planes, alignment of the X-ray beam to the reconstructed image and missed tissue; reproducibility of the tomosynthesis exposure; and homogeneity of the reconstructed focal planes.
Summaries of results from the tomosynthesis QC tests are presented together with some 2D results for comparison.
The tomosynthesis QC tests and analysis methods developed were successfully applied. The lessons learnt, which are detailed in the Discussion section, may be helpful to others embarking on DBT QC programmes.
DBT performance test equipment and analysis methods have been developed. The experience gained has contributed to the subsequent drafting of DBT QC protocols in the UK and Europe.
开发断层合成质量控制(QC)测试方法,并将其与既定的二维(2D)QC测试一起使用,以测量在与二维乳腺摄影的对比试验中使用的数字乳腺断层合成(DBT)系统的性能。
制定了DBT QC方案及相关分析方法,纳入了一些二维测试的改编以及一些新颖的测试。断层合成测试包括:标准乳腺模型的平均腺体剂量;重建焦平面的对比度噪声比;几何畸变;伪影扩散;重建焦平面的阈值对比度细节检测、X射线束与重建图像的对准以及漏诊组织;断层合成曝光的可重复性;以及重建焦平面的均匀性。
展示了断层合成QC测试结果的总结以及一些二维结果以供比较。
所开发的断层合成QC测试和分析方法得到了成功应用。在讨论部分详细阐述的经验教训可能对其他开展DBT QC计划的人有所帮助。
已开发出DBT性能测试设备和分析方法。所获得的经验有助于英国和欧洲后续起草DBT QC方案。