Neufang K F, Ewen K
Radiologisches Institut der Universität zu Köln.
Rontgenblatter. 1989 Apr;42(4):180-6.
The mutual dependence of various parameters--frame rate, radiation dose per frame, size of the image intensifier field, photon energy, frame mode, and patient diameter--is investigated systematically on a modern DSA equipment. The software programme of the DSA machine widely modifies these parameters according to the individual conditions, which may influence both radiation exposure and image quality. IV DSA requires low photon energies to exploit the low iodine signal. Higher frame rates, oblique views and small image intensifier fields should be used restrictively especially in the abdomen, as they tend to lower the radiation doses per frame and to increase photon energies, thus impairing contrast resolution. Full understanding of the physical and technical basics and familiarity with the software programme of own DSA machine are mandatory to allow appropriate selection of the various parameters, according to the individual patient and his/her diagnostic problems. They are the key to optimal imaging quality at a tolerable radiation exposure.
在一台现代数字减影血管造影(DSA)设备上,系统地研究了各种参数之间的相互依赖性,这些参数包括帧率、每帧辐射剂量、影像增强器视野大小、光子能量、帧模式以及患者直径。DSA机器的软件程序会根据个体情况广泛调整这些参数,这可能会影响辐射暴露和图像质量。静脉DSA需要低光子能量来利用低碘信号。特别是在腹部,应限制使用较高的帧率、斜位视图和较小的影像增强器视野,因为它们往往会降低每帧的辐射剂量并增加光子能量,从而损害对比分辨率。要根据个体患者及其诊断问题适当选择各种参数,必须全面了解物理和技术基础,并熟悉自己DSA机器的软件程序。它们是在可耐受的辐射暴露下实现最佳成像质量的关键。