Renninger Daniel, Ernst Arne, Todt Ingo
a Department of Otolaryngology , Head and Neck Surgery at Unfallkrankenhaus Berlin , Germany.
Acta Otolaryngol. 2016;136(3):241-4. doi: 10.3109/00016489.2015.1115552. Epub 2015 Dec 1.
Conclusion MRI examinations in patients with an alternatively coupled VSB can lead to unpleasant side-effects. However, the residual hearing was not impaired, whereas the hearing performance with the VSB was decreased in one patient which could be fixed by a surgical revision. Different experiences for the VSB 503 can be expected. Objective To investigate the in vivo effects of MRI scanning on the Vibrant Soundbridge system (VSB) with an alternatively coupled Floating Mass Transducer (FMT). Method Sixty-five VSB (502) implantees were included in this study. Of them, 42 questionnaires could be evaluated with the patients' statements about their medical, otological, and general condition before, during, and after an MRI scan which was indicated for different medical reasons, despite the previous implantation of an alternatively coupled Vibrant Soundbridge System. Results In four patients (9.5%), five MRI examinations were performed. These were done for different indications (e.g. knee and shoulder joint diagnostics). During the scanning, noise and subjectively perceived distortion of the implant were described. A deterioration of the hearing gain with the VSB in place was found in one patient. A decrease of the hearing threshold was not observed.
结论 对于采用交替耦合振动声桥(VSB)的患者,MRI检查可能会导致不良副作用。然而,残余听力并未受损,不过有1例患者佩戴VSB时的听力表现下降,这可通过手术修复来解决。VSB 503可能会有不同的情况。目的 研究MRI扫描对采用交替耦合浮动质量传感器(FMT)的振动声桥系统(VSB)的体内影响。方法 本研究纳入了65例植入VSB(502)的患者。其中,42份问卷可根据患者关于其在因不同医学原因进行MRI扫描之前、期间和之后的医疗、耳科及一般状况的陈述进行评估,尽管此前已植入交替耦合振动声桥系统。结果 4例患者(9.5%)进行了5次MRI检查。这些检查用于不同的指征(如膝关节和肩关节诊断)。在扫描过程中,描述了植入物的噪声和主观感觉到的失真情况。发现1例患者佩戴VSB时听力增益下降。未观察到听力阈值降低。