MatOrtho proximal interphalangeal joint arthroplasty: minimum 2-year follow-up.

作者信息

Flannery O, Harley O, Badge R, Birch A, Nuttall D, Trail I A

机构信息

Wrightington Hospital, Wigan, UK.

出版信息

J Hand Surg Eur Vol. 2016 Nov;41(9):910-916. doi: 10.1177/1753193415614251. Epub 2016 Jul 19.

Abstract

The MatOrtho proximal interphalangeal replacement is a cementless cobalt-chromium metal-on-polyethylene mobile-bearing surface replacement arthroplasty. The aim of this study is to report the outcome and complications of this implant at a minimum of 2 years follow-up from a single institution. A retrospective case review was performed on all MatOrtho proximal interphalangeal joint replacements performed with a minimum of 2 years follow-up. Patient demographics, diagnosis, implant revision and other surgical interventions were recorded. Subjective and objective outcomes were evaluated at latest follow-up, including pain scores, range of motion, function and radiographic assessment. A total of 109 implants were inserted in 56 patients. Nine implants (six patients) were lost to follow-up. Of the remaining 100 implants, 75 had been undertaken in females. The mean age at time of surgery was 64 years and the principal diagnosis was osteoarthritis in 74%. The mean follow-up was 47 months (range 24-77). Within the group there was a statistically significant diminution in pain. There was also an improvement in functional scores post-operatively. Improvement in range of motion was seen in those joints with a pre-operative range of motion greater than 20°. Radiologically there was no evidence of loosening or of implant subsidence at final follow-up. The revision rate was 13%. Nine joints were revised to the NeuFlex (silicone rubber) prosthesis, three were converted to an arthrodesis and one had exchange of the MatOrtho prosthesis. The survival of the MatOrtho proximal interphalangeal joint arthroplasty was 85% at a minimum of 2-years follow-up. Patients can be advised that the procedure achieves good pain relief, improvement in functional scores and may improve range of motion. We would, however, caution against this implant's use in joints that are either stiff or have significant deformity and/or instability pre-operatively.

摘要

MatOrtho近节指间关节置换术是一种非骨水泥型钴铬合金金属对聚乙烯活动轴承表面置换关节成形术。本研究的目的是报告来自单一机构、至少随访2年的该种植体的治疗结果及并发症。对所有进行了MatOrtho近节指间关节置换且至少随访2年的病例进行回顾性分析。记录患者的人口统计学资料、诊断、种植体翻修及其他手术干预情况。在最近一次随访时评估主观和客观结果,包括疼痛评分、活动范围、功能及影像学评估。56例患者共植入109枚种植体。9枚种植体(6例患者)失访。其余100枚种植体中,75枚植入女性患者体内。手术时的平均年龄为64岁,主要诊断为骨关节炎的占74%。平均随访时间为47个月(范围24 - 77个月)。组内疼痛有统计学意义的减轻。术后功能评分也有改善。术前活动范围大于20°的关节,其活动范围有所改善。影像学检查显示,末次随访时无松动或种植体下沉迹象。翻修率为13%。9个关节翻修为NeuFlex(硅橡胶)假体,3个关节改为关节融合术,1个关节更换了MatOrtho假体。MatOrtho近节指间关节成形术在至少随访2年时的生存率为85%。可以告知患者,该手术能有效缓解疼痛、改善功能评分,且可能改善活动范围。然而,对于术前僵硬或有明显畸形和/或不稳定的关节,我们建议谨慎使用该种植体。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验