The Potential of High Resolution Magnetic Resonance Microscopy in the Pathologic Analysis of Resected Breast and Lymph Tissue.

作者信息

Dashevsky Brittany Z, D'Alfonso Timothy, Sutton Elizabeth J, Giambrone Ashley, Aronowitz Eric, Morris Elizabeth A, Juluru Krishna, Ballon Douglas J

机构信息

Department of Radiology, Weill Cornell Medical College, Cornell University, New York, NY.

Department of Pathology, Weill Cornell Medical College, Cornell University, New York, NY.

出版信息

Sci Rep. 2015 Dec 7;5:17435. doi: 10.1038/srep17435.

Abstract

Pathologic evaluation of breast specimens requires a fixation and staining procedure of at least 12 hours duration, delaying diagnosis and post-operative planning. Here we introduce an MRI technique with a custom-designed radiofrequency resonator for imaging breast and lymph tissue with sufficient spatial resolution and speed to guide pathologic interpretation and offer value in clinical decision making. In this study, we demonstrate the ability to image breast and lymphatic tissue using 7.0 Tesla MRI, achieving a spatial resolution of 59 × 59 × 94 μm(3) with a signal-to-noise ratio of 15-20, in an imaging time of 56 to 70 minutes. These are the first MR images to reveal characteristic pathologic features of both benign and malignant breast and lymph tissue, some of which were discernible by blinded pathologists who had no prior training in high resolution MRI interpretation.

摘要

乳腺标本的病理评估需要至少持续12小时的固定和染色程序,这会延迟诊断和术后规划。在此,我们介绍一种采用定制设计射频谐振器的MRI技术,用于对乳腺和淋巴组织进行成像,具有足够的空间分辨率和速度,以指导病理解读并在临床决策中提供价值。在本研究中,我们展示了使用7.0特斯拉MRI对乳腺和淋巴组织进行成像的能力,在56至70分钟的成像时间内,实现了59×59×94μm³的空间分辨率,信噪比为15至20。这些是首批揭示良性和恶性乳腺及淋巴组织特征性病理特征的MR图像,其中一些特征对未接受过高分辨率MRI解读培训的盲法病理学家来说也是可辨别的。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5af1/4671009/01d2fa7034ac/srep17435-f1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验