医院中电子医疗设备安全使用的电磁场评估

Evaluation of Electromagnetic Fields in a Hospital for Safe Use of Electronic Medical Equipment.

作者信息

Ishida Kai, Fujioka Tomomi, Endo Tetsuo, Hosokawa Ren, Fujisaki Tetsushi, Yoshino Ryoji, Hirose Minoru

机构信息

Division of Healthcare Informatics, Faculty of Healthcare, Tokyo Healthcare University, Setagaya 3-11-3, Tokyo, 154-8568, Japan.

Advanced Technology Development Section, Building Technology Development Department, Technology Center, Taisei Corporation, Nase-cho 344-1, Yokohama, 245-0051, Japan.

出版信息

J Med Syst. 2016 Mar;40(3):46. doi: 10.1007/s10916-015-0411-3. Epub 2015 Dec 7.

Abstract

Establishment of electromagnetic compatibility is important in use of electronic medical equipment in hospitals. To evaluate the electromagnetic environment, the electric field intensity induced by electromagnetic radiation in broadcasting spectra coming from outside the hospital was measured in a new hospital building before any patients visited the hospital and 6 months after the opening of the hospital. Various incoming radio waves were detected on the upper floors, with no significant difference in measured levels before and after opening of the hospital. There were no cellphone terminal signals before the hospital opened, but these signals were strongly detected at 6 months thereafter. Cellphone base stations signals were strongly detected on the upper floors, but there were no signals at most locations in the basement and in the center of the building on the lower floors. A maximum electrical intensity of 0.28 V/m from cellphone base stations (2.1 GHz) was detected at the south end of the 2nd floor before the hospital opened. This value is lower than the EMC marginal value for general electronic medical equipment specified in IEC 60601-1-2 (3 V/m). Therefore, electromagnetic interference with electronic medical equipment is unlikely in this situation. However, cellphone terminal signals were frequently detected in non-base station signal areas. This is a concern, and understanding signal strength from cellphone base stations at a hospital is important for promotion of greater safety.

摘要

在医院使用电子医疗设备时,建立电磁兼容性很重要。为评估电磁环境,在新医院大楼有患者就诊之前以及医院开业6个月后,对来自医院外部广播频谱中的电磁辐射所感应的电场强度进行了测量。在较高楼层检测到了各种入射无线电波,医院开业前后测量水平没有显著差异。医院开业前没有手机终端信号,但此后6个月强烈检测到了这些信号。在较高楼层强烈检测到了手机基站信号,但在地下室以及较低楼层建筑物中心的大多数位置没有信号。医院开业前,在二楼南端检测到来自手机基站(2.1 GHz)的最大电场强度为0.28 V/m。该值低于IEC 60601-1-2规定的一般电子医疗设备的电磁兼容性临界值(3 V/m)。因此,在这种情况下不太可能对电子医疗设备产生电磁干扰。然而,在非基站信号区域经常检测到手机终端信号。这是一个问题,了解医院中手机基站的信号强度对于提高安全性很重要。