[新生儿脑血管血流速度的测量:一种连续多普勒超声检测方法]

[Measurement of blood flow velocity in the cerebral vessels of neonates: a method for continual doppler ultrasound detection].

作者信息

Zoban P, Karas J, Cerný M

出版信息

Cesk Pediatr. 1989 Jun;44(6):327-30.

PMID:2667766
Abstract

The authors refer in their paper to previous discussions on methodological problems associated with the application of Doppler continuous ultrasonography in neonatological practice. The methodological investigation was made in a group of twelve neonates with a mean birth weight of 2028 g and gestational age of 33.3 weeks. All were examined at the age of three weeks when the postpartum adaptation is already completed and the clinical condition is stabilized. After perfect mastering of the technique of measurement, the authors found that the mean error of assessment of PI on the internal carotid artery was 3.6%, on its siphon 2.7% and on the anterior cerebral artery 2.7%. The mean error of measurement for MVF on the same arteries was 7.7%, 13.7% and 6.1% resp. The greatest error was for EDFV on the internal carotid artery 11.7% on its siphon 15.8% and on the anterior cerebral artery 6.1%. The smallest mean error of assessment of all three parameters was on the anterior cerebral artery, the greatest error on the siphon of the internal carotid artery. On the investigated arteries no significant lateral difference was recorded. The mean values of PI, MFV and EDFV have a declining trend in the direction towards the periphery, i.e. in the direction from the internal carotid artery to the anterior cerebral artery. These trends were also confirmed in individual children. The authors assume that the method of continuous Doppler ultrasonography is despite limitations, inherent to its nature, suited for use in neonatological practice.

摘要

作者在其论文中提及了先前关于多普勒连续超声检查在新生儿科实践中应用的方法学问题的讨论。方法学研究是在一组12名新生儿中进行的,这些新生儿的平均出生体重为2028克,胎龄为33.3周。所有新生儿均在出生三周时接受检查,此时产后适应已完成,临床状况稳定。在熟练掌握测量技术后,作者发现,颈内动脉搏动指数(PI)评估的平均误差为3.6%,其虹吸部为2.7%,大脑前动脉为2.7%。同一动脉上平均血流速度(MVF)测量的平均误差分别为7.7%、13.7%和6.1%。颈内动脉舒张末期血流速度(EDFV)评估的最大误差为11.7%,其虹吸部为15.8%,大脑前动脉为6.1%。所有三个参数评估的最小平均误差出现在大脑前动脉,最大误差出现在颈内动脉虹吸部。在所研究的动脉上未记录到明显的侧别差异。PI、MFV和EDFV的平均值在向周边方向,即从颈内动脉到大脑前动脉的方向上呈下降趋势。这些趋势在个体儿童中也得到了证实。作者认为,尽管连续多普勒超声检查方法有其固有的局限性,但仍适用于新生儿科实践。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验