J Dent Hyg. 2015 Dec;89(6):384-9.
Technological advances in intra-oral receptors have resulted in film-holding devices that may or may not be interchangeable with photostimulable phosphor receptors. This study evaluated the number and types of technique errors that occurred when using PSP receptors with a standard film-holding device and a dual PSP/film-designed device.
The Rinn XCP-ORA® (Standard) and the Rinn Flip-Ray® PA device (Test) were compared using rectangular collimation. DenOptix® imaging plates (sizes 1 and 2) were used as receptors. Fourteen periapical (10-size 2 and 4-size 1) projections were exposed per full mouth series on each Dental X-ray Teaching and Training Replica with both devices. Five Dental X-ray Teaching and Training Replicas were exposed by 3 experienced radiographers. Data were analyzed using a paired t-test to determine differences in the performance scores between the 2 devices. Technique errors (receptor placement, vertical angulation, horizontal angulation and cone centering) were reported using frequencies. An experienced evaluator critiqued each projection.
A total of 15 full mouth series (210 projections) were taken per device. The mean performance scores per device were 88.4 (standard device) and 88.1 (test device) and were not statistically different (p=0.88). Cone centering errors were the most common error observed in both the standard (36%) and test (43%) devices. Receptor placement errors occurred when using the standard (12%) and test (9%) devices. Vertical and horizontal errors were <2% for both devices.
Devices designed for use with film may be used interchangeably with photostimulable phosphor receptors. Some difference was noticed between devices regarding error type and occurrence.
口腔内感受器的技术进步带来了胶片固定装置,这些装置可能与光激励荧光受体互换使用,也可能无法互换。本研究评估了使用光激励荧光受体与标准胶片固定装置和双光激励荧光/胶片设计装置时出现的技术错误的数量和类型。
使用矩形准直比较Rinn XCP-ORA®(标准)和Rinn Flip-Ray®全景片装置(测试)。使用DenOptix®成像板(尺寸1和2)作为感受器。在每个牙科X射线教学和培训模型上,用这两种装置对每个全口系列进行14次根尖片(10张尺寸2和4张尺寸1)投照。3名经验丰富的放射技师对5个牙科X射线教学和培训模型进行投照。使用配对t检验分析数据,以确定两种装置之间性能得分的差异。使用频率报告技术错误(感受器放置、垂直角度、水平角度和锥束中心)。一名经验丰富的评估人员对每个投照进行评判。
每个装置共拍摄了15个全口系列(210次投照)。每个装置的平均性能得分分别为88.4(标准装置)和88.1(测试装置),差异无统计学意义(p = 0.88)。锥束中心错误是在标准装置(36%)和测试装置(43%)中观察到的最常见错误。使用标准装置(12%)和测试装置(9%)时出现了感受器放置错误。两种装置的垂直和水平错误均<2%。
设计用于胶片的装置可与光激励荧光受体互换使用。在装置之间,关于错误类型和发生率存在一些差异。