Chreiteh Shadi S, Belhage Bo, Hoppe Karsten, Branebjerg Jens, Haahr Rasmus, Duun Sune, Thomsen Erik V
Annu Int Conf IEEE Eng Med Biol Soc. 2015;2015:6570-3. doi: 10.1109/EMBC.2015.7319898.
The respiratory rate (RR) is a clinically important vital sign and is a frequently used parameter in the general hospital wards. In current clinical practice, the monitoring of the RR is by manual count of the chest movement for one minute. This paper addresses a new approach where the respiratory rate is extracted using photoplethysmography (PPG) on the chest bone (sternum). Sternal PPG signals were acquired from 10 healthy subjects resting in a supine position. As reference signals, finger PPG, electrocardiogram (ECG), and capnography were simultaneously recorded during spontaneous and paced breathing. The sternal PPG signals were then compared with the reference signals in terms of Bland-Altman analysis, the power spectrum analysis and the magnitude squared coherence. The Bland-Altman analysis showed an average bias of 0.21 breaths/min between RR extracted from sternal PPG and capnography. The respiratory power content at the sternum was 78.8 (38) % in terms of the median and (the interquartile range). The cardiac content was 19 (18.4) % within the cardiac region. The results from the magnitude squared coherence analysis was 0.97 (0.09) in the respiratory region (6 to 27 breaths/min) and 0.98 (0.01) in the cardiac pulse region (30-120 beats/min). This preliminary study demonstrates the possibility of monitoring the RR from sternal PPG on a healthy group of subjects during rest.
呼吸频率(RR)是一项临床重要的生命体征,也是综合医院病房中常用的参数。在当前临床实践中,RR的监测是通过手动计数胸部运动一分钟来进行的。本文提出了一种新方法,即利用胸骨上的光电容积脉搏波描记法(PPG)提取呼吸频率。从10名仰卧休息的健康受试者身上采集胸骨PPG信号。作为参考信号,在自主呼吸和定频呼吸期间同时记录手指PPG、心电图(ECG)和二氧化碳图。然后,根据布兰德-奥特曼分析、功率谱分析和幅度平方相干性,将胸骨PPG信号与参考信号进行比较。布兰德-奥特曼分析显示,从胸骨PPG提取的RR与二氧化碳图之间的平均偏差为0.21次/分钟。胸骨处的呼吸功率含量中位数为78.8(38)%(四分位间距)。心脏区域内的心脏功率含量为19(18.4)%。幅度平方相干性分析结果在呼吸区域(6至27次/分钟)为0.97(0.09),在心脏搏动区域(30 - 120次/分钟)为0.98(0.01)。这项初步研究证明了在一组健康受试者休息期间从胸骨PPG监测RR的可能性。