Suppr超能文献

人工耳蜗电极不同插入技术对蜗内压力的影响

Effects of Different Insertion Techniques of a Cochlear Implant Electrode on the Intracochlear Pressure.

作者信息

Todt Ingo, Ernst Arneborg, Mittmann Philipp

机构信息

Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Berlin, Germany.

出版信息

Audiol Neurootol. 2016;21(1):30-7. doi: 10.1159/000442041. Epub 2016 Jan 22.

Abstract

To achieve a functional atraumatic insertion, low intracochlear pressure changes during the procedure are assumed to be important. The aim of this study was to observe intracochlear pressure changes due to different insertion techniques in a cochlear model. Cochlear implant electrode insertions were performed in an artificial cochlear model to record intracochlear pressure changes with a micropressure sensor to evaluate the maximum amplitude and frequency of pressure changes under different insertional conditions. We found statistically significant differences in the occurrence of intracochlear pressure peak changes comparing different techniques. Based on our model results, an insertion should be maximally supported to minimize micromovement-related pressure changes.

摘要

为实现功能性无创插入,术中较低的耳蜗内压力变化被认为很重要。本研究的目的是观察耳蜗模型中不同插入技术引起的耳蜗内压力变化。在人工耳蜗模型中进行人工耳蜗电极插入,用微压力传感器记录耳蜗内压力变化,以评估不同插入条件下压力变化的最大幅度和频率。比较不同技术时,我们发现耳蜗内压力峰值变化的发生率存在统计学显著差异。基于我们的模型结果,插入过程应得到最大程度的支撑,以尽量减少与微运动相关的压力变化。