Suppr超能文献

[心脏植入式电子设备家庭监测的应用经验]

[Experiences of applying home monitoring for cardiac implantable electronic devices].

作者信息

Qiu Liang, Xiang Meixiang, Wang Jian'an

机构信息

Department of Cardiology, Second Affiliated Hospital, Zhejiang University College of Medicine, Hangzhou 310009, China.

出版信息

Zhonghua Xin Xue Guan Bing Za Zhi. 2016 Jan;44(1):55-9. doi: 10.3760/cma.j.issn.0253-3758.2016.01.012.

Abstract

OBJECTIVE

To observe the results of remote home monitoring for cardiac implantable electronic devices (CIED).

METHODS

From November 2009 to July 2014, 37 patients implanted with home monitoring system CIED were enrolled, and 31 cases were implanted with dual chamber pacemakers, 2 with cardiac resynchronization therapy(CRT) and 4 with implantable cardioverter defibrillator (ICD). All patients received clinical routine follow-up.Data was automatic transmitted daily by remote home monitoring system, while special alarm events were transmitted at any time.The success rate of data transmission, abnormal events monitoring and pacing parameter changes were analyzed.

RESULTS

A total of 37 patients (24 males, mean age (70.0±9.6) years) were enrolled.Average follow-up time was (902±404) days.Success rate of data transmission was 97.6% (32 574), failed data transmission rate was 2.4% (801). Ninth-nine alerts were transmitted from 28 patients (75.7%), including 80 (80.8%) disease-related alerts and 19 (19.2%) system-related alerts.Average detection time of atrial fibrillation by remote home monitoring system to last clinical routine follow-up was 62 (19-120) days.There was no significant change in ventricular threshold, atrial and ventricle sensing during acute, subacute and chronic phases post the application of the steroid-eluting leads(all P>0.05).

CONCLUSIONS

The success rate of data transmission for CIED with remote home monitoring system is efficient, abnormal events can be timely detected.Meanwhile, remote home monitoring system also verified the stability of pacing parameters during acute, subacute and chronic phases post the application of the steroid-eluting leads.

摘要

目的

观察心脏植入式电子设备(CIED)远程家庭监测的结果。

方法

选取2009年11月至2014年7月期间植入家庭监测系统CIED的37例患者,其中植入双腔起搏器31例,心脏再同步治疗(CRT)2例,植入式心律转复除颤器(ICD)4例。所有患者均接受临床常规随访。数据由远程家庭监测系统每日自动传输,特殊报警事件随时传输。分析数据传输成功率、异常事件监测情况及起搏参数变化。

结果

共纳入37例患者(男性24例,平均年龄(70.0±9.6)岁)。平均随访时间为(902±404)天。数据传输成功率为97.6%(32 574次),数据传输失败率为2.4%(801次)。28例患者(75.7%)传输了99次警报,其中疾病相关警报80次(80.8%),系统相关警报19次(19.2%)。远程家庭监测系统检测到房颤至最后一次临床常规随访的平均时间为62(19 - 120)天。应用类固醇洗脱导线后急性、亚急性和慢性期心室阈值、心房和心室感知无明显变化(均P>0.05)。

结论

CIED远程家庭监测系统的数据传输成功率高,可及时检测到异常事件。同时,远程家庭监测系统也验证了应用类固醇洗脱导线后急性、亚急性和慢性期起搏参数的稳定性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验