Dimensional accuracy of digital dental models from cone-beam computed tomography scans of alginate impressions according to time elapsed after the impressions.

作者信息

Lee Sang-Mi, Hou Yanan, Cho Jin-Hyoung, Hwang Hyeon-Shik

机构信息

Postgraduate student, School of Dentistry, Chonnam National University, Gwangju, Korea.

Associate professor, Department of Orthodontics, Dental Science Research Institute, School of Dentistry, Chonnam National University, Gwangju, Korea.

出版信息

Am J Orthod Dentofacial Orthop. 2016 Feb;149(2):287-94. doi: 10.1016/j.ajodo.2015.08.014.

Abstract

INTRODUCTION

Our objective was to investigate the dimensional accuracy of digital dental models obtained from the cone-beam computed tomography (CBCT) scans of alginate impressions according to the time elapsed after the impressions were taken.

METHODS

Alginate impressions were obtained of 20 adults using 2 alginate materials: Alginoplast (Heraeus Kulzer, Hanau, Germany) and Cavex (Cavex Holland, Haarlem, The Netherlands). These impressions were stored in wet conditions and scanned by CBCT immediately after impression-taking and after storage times of 12, 24, 36, 48, 60, and 72 hours. After reconstruction of the 3-dimensional digital dental models, the models were measured, and the data were analyzed to determine the dimensional changes according to the time elapsed. The changes within the measurement error were regarded as clinically acceptable in this study.

RESULTS

The measurement errors ranged from 0.27 to 0.29 mm in the digital dental models obtained from CBCT scans of the alginate impressions. All measurements showed decreasing accuracy with greater elapsed time after the impressions were taken. Changes above the measurement error occurred at 24 and 36 hours for Cavex and Alginoplast, respectively.

CONCLUSIONS

Digital dental models can be obtained from CBCT scans of alginate impressions without sending them to a remote laboratory. However, the impressions need to be scanned within 24 hours; otherwise, dimensional changes lead to errors that exceed the error of measurement.

摘要

引言

我们的目的是根据藻酸盐印模采集后经过的时间,研究从藻酸盐印模的锥形束计算机断层扫描(CBCT)获得的数字牙科模型的尺寸精度。

方法

使用两种藻酸盐材料对20名成年人制取藻酸盐印模:藻酸盐塑料(德国哈瑙贺利氏古莎公司)和卡韦克斯(荷兰哈勒姆卡韦克斯公司)。这些印模保存在潮湿环境中,在印模制取后立即以及在12、24、36、48、60和72小时的储存时间后通过CBCT进行扫描。在重建三维数字牙科模型后,对模型进行测量,并分析数据以确定根据经过时间的尺寸变化。在本研究中,测量误差范围内的变化被视为临床上可接受的。

结果

从藻酸盐印模的CBCT扫描获得的数字牙科模型中的测量误差范围为0.27至0.29毫米。所有测量结果均显示,印模制取后经过的时间越长,精度越低。卡韦克斯和藻酸盐塑料分别在24小时和36小时出现超出测量误差的变化。

结论

无需将藻酸盐印模送到远程实验室,即可从其CBCT扫描中获得数字牙科模型。然而,印模需要在24小时内进行扫描;否则,尺寸变化会导致超出测量误差的误差。