藻酸盐印模的口内扫描仪和锥形束 CT 扫描所获数字模型的线性测量的准确性、可重复性和可靠性。

Validity, reliability, and reproducibility of linear measurements on digital models obtained from intraoral and cone-beam computed tomography scans of alginate impressions.

机构信息

Resident, Department of Orthodontics, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands.

出版信息

Am J Orthod Dentofacial Orthop. 2013 Jan;143(1):140-7. doi: 10.1016/j.ajodo.2012.06.018.

Abstract

INTRODUCTION

Digital 3-dimensional models are widely used for orthodontic diagnosis. The aim of this study was to assess the validity, reliability, and reproducibility of digital models obtained from the Lava Chairside Oral scanner (3M ESPE, Seefeld, Germany) and cone-beam computed tomography scans of alginate impressions for tooth-width measurements and the Bolton analysis.

METHODS

A digital model, an intraoral scan, and a plaster model were made for each of 22 subjects. Tooth-width measurements on the digital model and the intraoral scan were compared with those on the corresponding plaster models (gold standard). Intraclass correlation coefficients were calculated to determine the interexaminer reliability of the measurements of each method. The anterior and overall Bolton ratios were calculated for each participant and for each method. The paired t test was used to determine the validity. The scanning time for the intraoral scanner was registered and analyzed.

RESULTS

Tooth-width measurements of each tooth on the digital models and the intraoral scans did not differ significantly from those on the plaster models (P >0.05). The overall and anterior Bolton ratios from the 2 types of digital models differed significantly from the gold standard (P <0.05). However, the differences never exceeded 1.5 mm; this could be regarded as clinically insignificant. The scanning times of the intraoral scanner decreased significantly with the number of scans performed.

CONCLUSIONS

Both intraoral scanning and cone-beam computed tomography scanning of alginate impressions are valid, reliable, and reproducible methods to obtain dental measurements for diagnostic purposes.

摘要

简介

数字三维模型在正畸诊断中得到广泛应用。本研究旨在评估 Lava Chairside 口内扫描仪(3M ESPE,德国 Seefeld)获取的数字模型与藻酸盐印模的锥形束 CT 扫描在牙齿宽度测量和 Bolton 分析方面的有效性、可靠性和可重复性。

方法

对 22 名受试者的每个个体均制取数字模型、口内扫描和石膏模型。将数字模型和口内扫描上的牙齿宽度测量值与相应的石膏模型(金标准)进行比较。采用组内相关系数评估每种方法测量的组内可靠性。为每位参与者和每种方法计算前牙和整体 Bolton 比值。采用配对 t 检验确定有效性。记录并分析口内扫描仪的扫描时间。

结果

数字模型和口内扫描上各牙齿的牙齿宽度测量值与石膏模型相比无显著差异(P>0.05)。两种数字模型的整体和前牙 Bolton 比值与金标准相比有显著差异(P<0.05)。然而,差异从未超过 1.5mm,这可被视为临床无意义。口内扫描仪的扫描时间随扫描次数的增加而显著减少。

结论

口内扫描和锥形束 CT 扫描藻酸盐印模均是用于诊断目的获取牙齿测量值的有效、可靠和可重复的方法。