Suppr超能文献

迈向基于听觉稳态反应的术中听神经监测最佳范式。

Towards an optimal paradigm for intraoperative auditory nerve monitoring with auditory steady state responses.

作者信息

Rampp Stefan, Rensch Leonhard, Simmermacher Sebastian, Rahne Torsten, Strauss Christian, Prell Julian

机构信息

Department of Neurosurgery, University Hospital Halle (Saale), Ernst-Grube-Str. 40, 06097, Halle (Saale), Germany.

Department of Otorhinolaryngology, Head and Neck Surgery, University Hospital Halle (Saale), Ernst-Grube-Str. 40, 06097, Halle (Saale), Germany.

出版信息

J Clin Monit Comput. 2017 Feb;31(1):123-134. doi: 10.1007/s10877-016-9837-5. Epub 2016 Feb 6.

Abstract

Auditory steady state responses (ASSR) may offer an alternative to brainstem auditory evoked potentials for monitoring of the auditory nerve during surgical procedures. In the current study, we evaluated the influence of noise on ASSR characteristics in total intravenous anesthesia (TIVA). Simulated ASSR in real noise recorded during surgery under TIVA were constructed with known parameters. Influence of amplitude, modulation frequency, averaging sweeps and detection threshold on ASSR were evaluated. High amplitude, more sweeps and a liberal threshold facilitated detection. High amplitude ASSR (80 nV) were detected in up to 45 % with 16 s of data, in 80-90 % with 112 s. Near-threshold ASSR were detected in 0.8-25 %. False positives ranged between 0.3 and 10.3 %. Number of sweeps did not influence false positives. Amplitude errors varied between -61 and +39 % and improved with more averages but not with different thresholds. Modulation rate demonstrated the strongest influence on all parameters. 110 Hz yielded best, 90 Hz the worst results. Choice of parameters strongly influences detection and characteristics of ASSR. Optimal parameters enabled detection after 16 s in 45 %. Due to specific noise characteristics, modulation has a critical impact, which is currently not sufficiently recognized in ASSR studies.

摘要

听觉稳态反应(ASSR)可能为手术过程中监测听神经提供一种替代脑干听觉诱发电位的方法。在本研究中,我们评估了噪声对全静脉麻醉(TIVA)期间ASSR特征的影响。利用已知参数构建了TIVA下手术期间记录的真实噪声中的模拟ASSR。评估了幅度、调制频率、平均扫描次数和检测阈值对ASSR的影响。高幅度、更多扫描次数和宽松的阈值有助于检测。在长达16秒的数据中,高达45%的情况下检测到高幅度ASSR(80 nV),在112秒时为80 - 90%。接近阈值的ASSR检测率为0.8 - 25%。假阳性率在0.3%至10.3%之间。扫描次数不影响假阳性。幅度误差在 - 61%至 + 39%之间变化,更多平均次数可改善误差,但不同阈值则不能。调制率对所有参数的影响最为显著。110 Hz产生的结果最佳,90 Hz最差。参数的选择对ASSR的检测和特征有强烈影响。最佳参数能在16秒后使45%的情况下检测到ASSR。由于特定的噪声特征,调制具有关键影响,但目前在ASSR研究中尚未得到充分认识。