Normal ranges and test-retest reproducibility of flow and velocity parameters in intracranial arteries measured with phase-contrast magnetic resonance imaging.

作者信息

Correia de Verdier Maria, Wikström Johan

机构信息

Department of Radiology, Uppsala University, Uppsala University Hospital, 751 85, Uppsala, Sweden.

出版信息

Neuroradiology. 2016 May;58(5):521-31. doi: 10.1007/s00234-016-1661-6. Epub 2016 Feb 16.

Abstract

INTRODUCTION

The purpose of the present study was to investigate normal ranges and test-retest reproducibility of phase-contrast MRI (PC-MRI)-measured flow and velocity parameters in intracranial arteries.

METHODS

Highest flow (HF), lowest flow (LF), peak systolic velocity (PSV), and end diastolic velocity (EDV) were measured at two dates in the anterior (ACA), middle (MCA), and posterior (PCA) cerebral arteries of 30 healthy volunteers using two-dimensional PC-MRI at 3 T. Least detectable difference (LDD) was calculated.

RESULTS

In the left ACA, HF was (mean (range, LDD)) 126 ml/min (36-312, 59 %), LF 61 ml/min (0-156, 101 %), PSV 64 cm/s (32-141, 67 %), and EDV 35 cm/s (18-55, 42 %); in the right ACA, HF was 154 ml/min (42-246, 49 %), LF 77 ml/min (0-156, 131 %), PSV 75 cm/s (26-161, 82 %), and EDV 39 cm/s (7-59, 67 %). In the left MCA, HF was 235 ml/min (126-372, 35 %), LF 116 ml/min (42-186, 48 %), PSV 90 cm/s (55-183, 39 %), and EDV 46 cm/s (20-66, 28 %); in the right MCA, HF was 238 ml/min (162-342, 44 %), LF 120 ml/min (72-216, 48 %), PSV 88 cm/s (55-141, 35 %), and EDV 45 cm/s (26-67, 23 %). In the left PCA, HF was 108 ml/min (42-168, 54 %), LF 53 ml/min (18-108, 64 %), PSV 50 cm/s (24-77, 63 %), and EDV 28 cm/s (14-40, 45 %); in the right PCA, HF was 98 ml/min (30-162, 49 %), LF 49 ml/min (12-84, 55 %), PSV 47 cm/s (27-88, 59 %), and EDV 27 cm/s (16-41, 45 %).

CONCLUSION

PC-MRI-measured flow and velocity parameters in the main intracranial arteries have large normal ranges. Reproducibility is highest in MCA.

摘要

引言

本研究旨在探讨相位对比磁共振成像(PC-MRI)测量颅内动脉血流和速度参数的正常范围及重测信度。

方法

使用3T的二维PC-MRI在两个时间点测量30名健康志愿者的前交通动脉(ACA)、大脑中动脉(MCA)和大脑后动脉(PCA)的最高血流(HF)、最低血流(LF)、收缩期峰值流速(PSV)和舒张末期流速(EDV)。计算最小可检测差异(LDD)。

结果

在左侧ACA中,HF为(均值(范围,LDD))126毫升/分钟(36 - 312,59%),LF为61毫升/分钟(0 - 156,101%),PSV为64厘米/秒(32 - 141,67%),EDV为35厘米/秒(18 - 55,42%);在右侧ACA中,HF为154毫升/分钟(42 - 246,49%),LF为77毫升/分钟(0 - 156,131%),PSV为75厘米/秒(26 - 161,82%),EDV为39厘米/秒(7 - 59,67%)。在左侧MCA中,HF为235毫升/分钟(126 - 372,35%),LF为116毫升/分钟(42 - 186,48%),PSV为90厘米/秒(55 - 183,39%),EDV为46厘米/秒(20 - 66,28%);在右侧MCA中,HF为238毫升/分钟(162 - 342,44%),LF为120毫升/分钟(72 - 216,48%),PSV为88厘米/秒(55 - 141,35%),EDV为45厘米/秒(26 - 67,23%)。在左侧PCA中,HF为108毫升/分钟(42 - 168,54%),LF为53毫升/分钟(18 - 108,64%),PSV为50厘米/秒(24 - 77,63%),EDV为28厘米/秒(14 - 40,45%);在右侧PCA中,HF为98毫升/分钟(30 - 162,49%),LF为49毫升/分钟(12 - 84,55%),PSV为47厘米/秒(27 - 88,59%),EDV为27厘米/秒(16 - 41,45%)。

结论

PC-MRI测量的主要颅内动脉血流和速度参数具有较大的正常范围。MCA的重测信度最高。