Suppr超能文献

使用27G双占空比玻璃体切割器治疗下沉的晶状体核。

Treatment of Dropped Nucleus with a 27-Gauge Twin Duty Cycle Vitreous Cutter.

作者信息

Watanabe Akira, Tsuzuki Akane, Arai Kota, Gekka Tamaki, Tsuneoka Hiroshi

机构信息

Department of Ophthalmology, The Jikei University School of Medicine, Tokyo, Japan.

出版信息

Case Rep Ophthalmol. 2016 Jan 14;7(1):44-8. doi: 10.1159/000443721. eCollection 2016 Jan-Apr.

Abstract

We report herein a method for the treatment of dropped nucleus during cataract surgery with a 27-gauge twin duty cycle (TDC) vitreous cutter. When a TDC vitreous cutter is used, suction flow volume is maintained even when the cutter is driven at a high speed. This enables an Emery-Little grade 3 nucleus that had been difficult to treat with a conventional 27-gauge cutter to be successfully excised using only a vitreous cutter, with no intra- or postoperative complications. A dropped lens during cataract surgery of up to moderate hardness can be removed using a TDC cutter alone with a 27-gauge cutter system.

摘要

我们在此报告一种使用27号双工作周期(TDC)玻璃体切割器治疗白内障手术中晶状体核掉落的方法。当使用TDC玻璃体切割器时,即使切割器高速驱动,抽吸流量也能保持稳定。这使得以往难以用传统27号切割器处理的埃默里-利特尔3级核,仅使用玻璃体切割器就能成功切除,且术中及术后均无并发症。对于白内障手术中掉落的硬度适中的晶状体,仅使用27号切割器系统的TDC切割器即可将其取出。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0794/4748802/d51a475aedc3/cop-0007-0044-g01.jpg

相似文献

1
Treatment of Dropped Nucleus with a 27-Gauge Twin Duty Cycle Vitreous Cutter.
Case Rep Ophthalmol. 2016 Jan 14;7(1):44-8. doi: 10.1159/000443721. eCollection 2016 Jan-Apr.
2
A 27-gauge instrument system for transconjunctival sutureless microincision vitrectomy surgery.
Ophthalmology. 2010 Jan;117(1):93-102.e2. doi: 10.1016/j.ophtha.2009.06.043. Epub 2009 Oct 31.
3
Experimental Assessment of the Performance of Dual Pneumatic Vitreous Cutters According to Gauge and Cut Rate.
Korean J Ophthalmol. 2023 Aug;37(4):307-313. doi: 10.3341/kjo.2023.0045. Epub 2023 Jul 3.
4
Performance analysis of new-generation vitreous cutters.
Graefes Arch Clin Exp Ophthalmol. 2008 Jan;246(1):61-7. doi: 10.1007/s00417-007-0672-8. Epub 2007 Sep 18.
5
Primary posterior capsulotomy using a 25-gauge vitreous cutter in vitrectomy combined with cataract surgery.
J Cataract Refract Surg. 2010 Jan;36(1):2-5. doi: 10.1016/j.jcrs.2009.07.049.
6
Intraoperative breakage of a 25-gauge vitreous cutter.
Am J Ophthalmol. 2004 Nov;138(5):867-9. doi: 10.1016/j.ajo.2004.05.035.
8
Performance analysis of millennium vitreous enhancer™ system.
Ophthalmic Surg Lasers Imaging. 2011 Mar-Apr;42(2):162-7. doi: 10.3928/15428877-20101223-03. Epub 2010 Dec 30.
9
Performance analysis of ultrahigh-speed vitreous cutter system.
Retina. 2013 May;33(5):928-32. doi: 10.1097/IAE.0b013e31826f069e.

引用本文的文献

1
Evaluation of the Efficiency and Safety of a 27-Gauge 20,000 Cuts per Minute Vitreous Cutter.
Clin Ophthalmol. 2023 Jul 18;17:2037-2043. doi: 10.2147/OPTH.S418371. eCollection 2023.
3
Impact of different manufacturers and gauge sizes on the performance of backflush needle.
Sci Rep. 2020 Dec 8;10(1):21452. doi: 10.1038/s41598-020-78668-6.
4
Comparative study of 27-gauge and 25-gauge vitrectomy performed as day surgery.
Int Ophthalmol. 2018 Aug;38(4):1575-1582. doi: 10.1007/s10792-017-0625-0. Epub 2017 Jul 3.
5
Review of Small Gauge Vitrectomy: Progress and Innovations.
J Ophthalmol. 2017;2017:6285869. doi: 10.1155/2017/6285869. Epub 2017 May 10.
6
Vitreoretinal instruments: vitrectomy cutters, endoillumination and wide-angle viewing systems.
Int J Retina Vitreous. 2016 Dec 5;2:28. doi: 10.1186/s40942-016-0052-9. eCollection 2016.

本文引用的文献

1
Treatment of a dislocated lens by transcorneal vitrectomy and bimanual phacoemulsification.
Clin Ophthalmol. 2014 Aug 18;8:1539-42. doi: 10.2147/OPTH.S66492. eCollection 2014.
2
Phacoemulsification with perfluorocarbon liquid using a 23-gauge transconjunctival sutureless vitrectomy for the management of dislocated crystalline lenses.
Graefes Arch Clin Exp Ophthalmol. 2013 May;251(5):1267-72. doi: 10.1007/s00417-012-2170-x. Epub 2012 Oct 7.
3
23-Gauge transconjunctival pars plana vitrectomy for removal of retained lens fragments.
Am J Ophthalmol. 2011 Oct;152(4):624-7. doi: 10.1016/j.ajo.2011.04.003. Epub 2011 Jul 2.
4
Management of retained lens fragments in complicated cataract surgery.
Curr Opin Ophthalmol. 2011 Mar;22(2):137-40. doi: 10.1097/ICU.0b013e3283436fc5.
5
A new dual port cutter system for vitrectomy surgery.
Retina. 2010 Oct;30(9):1515-9. doi: 10.1097/IAE.0b013e3181ea48f9.
6
A 27-gauge instrument system for transconjunctival sutureless microincision vitrectomy surgery.
Ophthalmology. 2010 Jan;117(1):93-102.e2. doi: 10.1016/j.ophtha.2009.06.043. Epub 2009 Oct 31.
7
Clinical predictors and outcomes of pars plana vitrectomy for retained lens material after cataract extraction.
Am J Ophthalmol. 2009 Apr;147(4):587-594.e1. doi: 10.1016/j.ajo.2008.10.026. Epub 2009 Feb 4.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验