Choudhary Pratik, Olsen Birthe S, Conget Ignacio, Welsh John B, Vorrink Linda, Shin John J
1 King's College London , London, United Kingdom .
2 Herlev Hospital , Herlev, Denmark .
Diabetes Technol Ther. 2016 May;18(5):288-91. doi: 10.1089/dia.2015.0324. Epub 2016 Feb 23.
The MiniMed 640G sensor-augmented insulin pump system (Medtronic, Inc., Northridge, CA) can automatically suspend insulin delivery in advance of predicted hypoglycemia and restart it upon recovery. The aims of this analysis were to determine the rate at which predicted hypoglycemia was avoided with this strategy, as well as to assess user acceptance of the system and its insulin management features.
Forty subjects with type 1 diabetes used the system for 4 weeks. We retrospectively evaluated performance of the system, using downloaded pump and sensor data, and evaluated user acceptance via questionnaires.
There were 2,322 suspend before low events (2.1 per subject-day). The mean (± SD) duration of pump suspension events was 56.4 ± 9.6 min, and the mean subsequent sensor glucose (SG) nadir was 71.8 ± 5.2 mg/dL. SG values following 1,930 (83.1%) of the predictive suspensions did not reach the preset low limit. Nadir SG values of ≤50 and ≤60 mg/dL were seen in 207 (8.9%) and 356 (15.3%) of the predictive suspensions, respectively. Blood glucose (BG) and SG values before and during the study were comparable (P > 0.05). The mean absolute relative difference between paired SG and BG values was 10.9 ± 13.8%. Subjects felt confident using the system, agreed that it helped protect them from hypoglycemia, and wished to continue using it.
Automatic insulin pump suspension as implemented in the MiniMed 640G system can help patients avoid hypoglycemia, without significantly increasing hyperglycemia.
美敦力640G传感器增强型胰岛素泵系统(美敦力公司,加利福尼亚州北岭)可在预测低血糖发生前自动暂停胰岛素输注,并在恢复后重新启动。本分析的目的是确定该策略避免预测低血糖的发生率,以及评估用户对该系统及其胰岛素管理功能的接受度。
40例1型糖尿病患者使用该系统4周。我们通过下载的泵和传感器数据回顾性评估系统性能,并通过问卷调查评估用户接受度。
共有2322次低血糖前暂停事件(每位受试者每天2.1次)。泵暂停事件的平均(±标准差)持续时间为56.4±9.6分钟,随后的平均传感器葡萄糖(SG)最低点为71.8±5.2mg/dL。1930次(83.1%)预测性暂停后的SG值未达到预设的低限。在207次(8.9%)和356次(15.3%)预测性暂停中,SG最低点值分别≤50mg/dL和≤60mg/dL。研究前后的血糖(BG)和SG值具有可比性(P>0.05)。配对的SG和BG值之间的平均绝对相对差异为10.9±13.8%。受试者对使用该系统感到有信心,认为它有助于保护他们免受低血糖影响,并希望继续使用。
美敦力640G系统实施的自动胰岛素泵暂停可帮助患者避免低血糖,而不会显著增加高血糖。