Radiother Oncol. 2016 Mar;118(3):521-7. doi: 10.1016/j.radonc.2016.02.019. Epub 2016 Feb 24.
To investigate the hypothesis that CT ventilation functional image-based IMRT plans designed to avoid irradiating highly-functional lung regions are comparable to single-photon emission CT (SPECT) ventilation functional image-based plans.
Three IMRT plans were created for eight thoracic cancer patients using: (1) CT ventilation functional images, (2) SPECT ventilation functional images, and (3) anatomic images (no functional images). CT ventilation images were created by deformable image registration of 4D-CT image data sets and quantitative analysis. The resulting plans were analyzed for the relationship between the deviations of CT-functional plan metrics from anatomic plan metrics (ΔCT-anatomic) and those of SPECT-functional plans (ΔSPECT-anatomic), and moreover for agreements of various metrics between the CT-functional and SPECT-functional plans.
The relationship between ΔCT-anatomic and ΔSPECT-anatomic was strong (e.g., R=0.94; linear regression slope 0.71). The average differences and 95% limits of agreement between the CT-functional and SPECT-functional plan metrics (except for monitor units) for various structures were mostly less than 1% and 2%, respectively.
This study demonstrated a reasonable agreement between the CT ventilation functional image-based IMRT plans and SPECT-functional plans, suggesting the potential for CT ventilation imaging to serve as a surrogate for SPECT ventilation in functional image-guided radiotherapy.