Acute transient non-physiological over-sensing in the ventricle with a DF4 lead.

作者信息

Ng Kevin Kit, Gould Paul A

机构信息

Princess Alexandra Hospital, Department of Cardiology, Queensland, Australia.

Princess Alexandra Hospital, Department of Cardiology, University of Queensland, Department of Medicine, Brisbane, Australia.

出版信息

Indian Pacing Electrophysiol J. 2015 Oct 19;15(4):216-9. doi: 10.1016/j.ipej.2015.10.008. eCollection 2015 Jul-Aug.

Abstract

The DF-4 is a new defibrillator lead technology. We present two cases of non-physiological transient ventricular over-sensing in patients who underwent implantation of an ICD for secondary prevention. Case 1 had ventricular over-sensing during pacing threshold evaluation post defibrillation testing while Case 2 had the lead integrity alert triggered immediately post discharge with transient over-sensing. No lead-connector issues were found. Case 1 was likely due to improper venting of the header and trapped air. Case 2 was hypothesized to be due to intermittent header pin non-contact secondary to blood in the header. These cases reveal that DF-4 leads are subject to both reported and potentially novel causes of transient acute ventricular over-sensing.

摘要

DF-4是一种新型除颤器导线技术。我们报告两例接受植入式心律转复除颤器(ICD)二级预防的患者出现非生理性短暂心室过度感知的病例。病例1在除颤测试后的起搏阈值评估期间出现心室过度感知,而病例2在出院后立即触发导线完整性警报并伴有短暂过度感知。未发现导线连接器问题。病例1可能是由于接头排气不当和空气滞留所致。病例2据推测是由于接头中有血液导致接头引脚间歇性非接触所致。这些病例表明,DF-4导线会出现已报道的以及潜在的新型短暂急性心室过度感知原因。