Schwahofer Andrea, Feist Harald, Georg Holger, Häring Peter, Schlegel Wolfgang
Department of Radiation Therapy, Vivantes Clinic Neukoelln, Berlin, Germany; Medical Physics in Radiation Therapy, German Cancer Research Center, Heidelberg, Germany.
Department of Radiation Therapy, Ludwig-Maximilians-University (LMU), Munich, Germany.
Z Med Phys. 2017 Mar;27(1):13-20. doi: 10.1016/j.zemedi.2016.02.003. Epub 2016 Mar 11.
The aim of this study has been the experimental determination of the energy dependent dose-to-water response of TLD600 and TLD700 thermoluminescent detectors (Harshaw) in X-ray beams with mean photon energies from about 20 to 200keV in comparison with Co gamma rays and 6MV X-rays. Experiments were carried out in collaboration with the German secondary standard laboratory PTW Freiburg. The energy dependent relative responses of TLD600 and TLD700 thermoluminescence detectors were determined at radiation qualities between 30kV and 280kV. The overall uncertainty of the measured values was characterized by standard deviations varying from 1.2 to 3%. The present results agree with previous studies on the energy dependent dose-to-water response of TLD100. As an application example, the results were used to measure doses associated with X-ray imaging in image-guided radiotherapy.