Suppr超能文献

CT立体定向装置在脑病变活检中的临床应用

[Clinical application of CT stereotactic device in biopsy of brain lesions].

作者信息

Liu X G

出版信息

Zhonghua Fang She Xue Za Zhi. 1989 Oct;23(5):257-60.

PMID:2697515
Abstract

A new CT stereotactic device named version WL-86 was designed and manufactured by the authors. Phantom puncture and cadaver puncture tests showed that its accuracy is less than 1 mm. We have also made a new type of biopsy needle--the rotary and translational cutting biopsy needle. A definite pathological diagnosis was made in 93.7% (30/32) of patients by this technique: The practicability of the device, the indications and clinical value of CT-guided stereotactic biopsy were discussed.

摘要

作者设计并制造了一种名为WL - 86型的新型CT立体定向装置。模型穿刺和尸体穿刺测试表明其精度小于1毫米。我们还制作了一种新型活检针——旋转平移切割活检针。通过该技术,93.7%(32例中的30例)的患者获得了明确的病理诊断:讨论了该装置的实用性、CT引导立体定向活检的适应证及临床价值。