Albogha Mhd Hassan, Kitahara Toru, Todo Mitsugu, Hyakutake Hiroto, Takahashi Ichiro
Section of Orthodontics and Dentofacial Orthopedics, Faculty of Dental Science, Kyushu University, 3-1-1, Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan.
Research Institute of Applied Mechanics, Kyushu University, Kasuga, 816-8580, Japan.
Ann Biomed Eng. 2016 Oct;44(10):2948-2956. doi: 10.1007/s10439-016-1584-8. Epub 2016 Mar 16.
Factors responsible for the success or failure of orthodontic mini-implants (OMIs) in clinical settings are unclear. Failure of OMIs was found to be associated with increased maximum principal strain (MaxPN) when assessed using the subject-specific finite element (FE) modeling technique. The purpose of the present study was to identify factors that increase MaxPN and thereby predispose the OMI to failure. Using the FE method, MaxPN was calculated around 28 OMIs placed in orthodontic patients, 6 of which failed during the first 5 months. Sixteen potential risk factors related to patients or to OMI position were measured on computerized tomographic images or calculated in FE models. The impact of these factors on MaxPN was verified using regression analysis. Three factors were found to have significant nonlinear relationships with MaxPN: cortical bone quality, vertical angulation of the OMI, and proximity of the OMI to the tooth in the direction of force. In conclusion, failure of an OMI is a multifactorial problem, and position and angulation of the implant are among the affecting factors. Slight apical inclination and positioning at least 1 mm off the root in the direction of force may significantly decrease failure probability.
正畸微型种植体(OMIs)在临床环境中成功或失败的相关因素尚不清楚。当使用个体化有限元(FE)建模技术进行评估时,发现OMIs的失败与最大主应变(MaxPN)增加有关。本研究的目的是确定增加MaxPN从而使OMI易于失败的因素。使用有限元方法,计算了正畸患者植入的28个OMIs周围的MaxPN,其中6个在前5个月内失败。在计算机断层扫描图像上测量或在有限元模型中计算了与患者或OMI位置相关的16个潜在风险因素。使用回归分析验证了这些因素对MaxPN的影响。发现三个因素与MaxPN有显著的非线性关系:皮质骨质量、OMI的垂直角度以及OMI在力的方向上与牙齿的接近程度。总之,OMI的失败是一个多因素问题,种植体的位置和角度是影响因素之一。轻微的根尖倾斜以及在力的方向上距离牙根至少1毫米的定位可能会显著降低失败概率。