Arenkov P Ia, Zhmareva E N, Berezin V A
Ukr Biokhim Zh (1978). 1989 Nov-Dec;61(6):89-92.
The enzyme-linked immunoassay modification has been worked out. The method combines advantages of membrane technology of antigen immobilization which is used in the enzyme immunosensory technique and of conventional enzyme-linked immunosorbent assay. The nitrocellulose and polypropylene membranes are used as a solid-phase. The purified rabbit immunoglobulin G is immobilized on the surface of membranes as the first layer. The competitive immunoassay is employed. The immunoglobulin G concentration range is 1-1000 ng/ml. The membranes with the immobilized antigen can be repeatedly used after incubation in 0.1 M glycine buffer, pH 2.5. The dry membrane with the immobilized antigen can be used after keeping for 6 months in refrigerator at 4 degrees C without changing the concentration range measured.