Herrnsdorf L, Petersson H
Medical Radiation Physics Malmö, Department of Translational Medicine, Lund University, Skåne University Hospital Malmö, 205 02 Malmö, Sweden
Medical Radiation Physics Malmö, Department of Translational Medicine, Lund University, Skåne University Hospital Malmö, 205 02 Malmö, Sweden.
Radiat Prot Dosimetry. 2016 Jun;169(1-4):347-52. doi: 10.1093/rpd/ncw042. Epub 2016 Mar 17.
Kerma-area product (KAP) meters have a pronounced energy dependence when measuring air KAP for lightly filtered X-ray beams (RQR). Today, it is also common with more heavily filtered beams. In this work, the energy dependence for lightly as well as heavily filtered beams (RQC) was investigated for several KAP meter models. The relative energy dependence of the readings of an external and an internal KAP meter was determined relative to an ionisation chamber, which had been calibrated at the primary standards laboratory. As a complement to the measurements, the sensitivity of a KAP meter for various X-ray beam qualities was modelled using Monte Carlo simulations of photon transport and absorption. The result showed a variation in relative energy dependence of up to 30 % for KAP meters for RQC beam qualities compared with RQR qualities. A reduced sensitivity of KAP meters for heavily filtered beams in comparison with lightly filtered ones was found, and it is important that the beam-specific radiation quality correction factors are applied to correct the registered KAP values.
比释动能面积乘积(KAP)仪在测量经轻度过滤的X射线束(RQR)的空气KAP时具有明显的能量依赖性。如今,使用过滤程度更高的射线束也很常见。在这项工作中,针对几种KAP仪型号研究了轻度和重度过滤射线束(RQC)的能量依赖性。相对于已在基准标准实验室校准的电离室,确定了外部和内部KAP仪读数的相对能量依赖性。作为测量的补充,使用光子传输和吸收的蒙特卡罗模拟对KAP仪对各种X射线束质量的灵敏度进行了建模。结果表明,与RQR射线质量相比,RQC射线质量的KAP仪的相对能量依赖性变化高达30%。发现KAP仪对重度过滤射线束的灵敏度低于轻度过滤射线束,并且应用特定于射线束的辐射质量校正因子来校正记录的KAP值非常重要。