Ayyildiz Onder, Kucukevcilioglu Murat, Ozge Gokhan, Koylu Mehmet Talay, Ozgonul Cem, Gokce Gokcen, Mumcuoglu Tarkan, Durukan Ali Hakan, Mutlu Fatih Mehmet
a Department of Ophthalmology , GATA Medical School , Ankara , Turkey.
b Department of Ophthalmology , Tatvan Military Hospital , Tatvan , Turkey.
Postgrad Med. 2016 May;128(4):439-43. doi: 10.1080/00325481.2016.1171119. Epub 2016 Apr 8.
To compare peripapillary choroidal thickness (PP-CT) measurements using a spectral domain optical coherence tomography (SD-OCT) device with and without enhanced depth imaging (EDI).
Sixty healthy subjects aged from 18 to 40 years were included in this study. PP-CTs were measured in the right eyes by manual segmentation via SD-OCT both with and without EDI. The intraclass correlation coefficient (ICC) for each technique and comparison of PP-CT measurements between two techniques were evaluated. The correlation between retinal nerve fiber layer (RNFL) thickness and PP-CT was also explored on images of SD-OCT without EDI.
The PP-CT measurements of 55 subjects were evaluated. The ICC was 0.999 (95% CI: 0.998-1.0, p < 0.001) for SD-OCT with EDI and 0.996 (95% CI: 0.995-0.997, p < 0.001) for SD-OCT without EDI. The mean PP-CT measurements in all regions and the overall mean PP-CT measurements between the two techniques were not different (p > 0.05). Additionally, there was no correlation between RNFL thickness and PP-CT (r = -0.109; p = 0.335).
The PP-CT measurements via SD-OCT without EDI were consistent with the measurements via SD-OCT with EDI. Ophthalmologists who do not have access to EDI technology can use images of SD-OCT without EDI to measure the peripapillary choroid for research purposes. However, thicker peripapillary choroids cannot be measured using this technique and require further modifications or newer technologies, such as SD-OCT with EDI.
比较使用具有和不具有增强深度成像(EDI)的光谱域光学相干断层扫描(SD-OCT)设备测量视乳头周围脉络膜厚度(PP-CT)的情况。
本研究纳入了60名年龄在18至40岁之间的健康受试者。通过SD-OCT分别在有和没有EDI的情况下,通过手动分割测量右眼的PP-CT。评估了每种技术的组内相关系数(ICC)以及两种技术之间PP-CT测量值的比较。还在没有EDI的SD-OCT图像上探讨了视网膜神经纤维层(RNFL)厚度与PP-CT之间的相关性。
对55名受试者的PP-CT测量值进行了评估。具有EDI的SD-OCT的ICC为0.999(95%CI:0.998 - 1.0,p < 0.001),没有EDI的SD-OCT的ICC为0.996(95%CI:0.995 - 0.997,p < 0.001)。两种技术在所有区域的平均PP-CT测量值以及总体平均PP-CT测量值没有差异(p > 0.05)。此外,RNFL厚度与PP-CT之间没有相关性(r = -0.109;p = 0.335)。
通过没有EDI的SD-OCT测量的PP-CT与通过有EDI的SD-OCT测量的结果一致。无法使用EDI技术的眼科医生可以使用没有EDI的SD-OCT图像来测量视乳头周围脉络膜以用于研究目的。然而,使用该技术无法测量较厚的视乳头周围脉络膜,需要进一步改进或采用更新的技术,如有EDI的SD-OCT。