Wu Meng-Ai, Xu Wei-Xin, Lyu Zhe, Shen Li-Jun
Eye Hospital of Wenzhou Medical University, Wenzhou 325027, Zhejiang Province, China.
Int J Ophthalmol. 2017 Oct 18;10(10):1539-1544. doi: 10.18240/ijo.2017.10.10. eCollection 2017.
To assess peripapillary retinal nerve fiber layer (RNFL) and choroidal thickness obtained with enhanced depth imaging (EDI) mode compared with those obtained without EDI mode using Heidelberg Spectralis optical coherence tomography (OCT).
Fifty eyes of 25 normal healthy subjects and 32 eyes of 20 patients with different eye diseases were included in the study. All subjects underwent 3.4 mm diameter peripapillary circular OCT scan centered on the optic disc using both the conventional and the EDI OCT protocols. The visualization of RNFL and choroidoscleral junction was assessed using an ordinal scoring scale. The paired -test, intraclass correlation coefficient (ICC), 95% limits of agreement (LoA), and Bland and Altman plots were used to test the agreement of measurements.
The visibility score of RNFL obtained with and without EDI was of no significant difference (=0.532), the visualization of choroidoscleral junction was better using EDI protocol than conventional protocol (<0.001). Peripapillary RNFL thickness obtained with EDI was slightly thicker than that obtained without EDI (103.25±9.42 µm 101.87±8.78 µm, =0.010). The ICC of the two protocols was excellent with the value of 0.867 to 0.924, the 95% LoA of global RNFL thickness was between -10.0 to 7.4 µm. Peripapillary choroidal thickness obtained with EDI was slightly thinner than that obtained without EDI (147.23±51.04 µm 150.90±51.84 µm, <0.001). The ICC was also excellent with the value of 0.960 to 0.987, the 95% LoA of global choroidal thickness was between -12.5 to 19.8 µm.
Peripapillary circular OCT scan with or without EDI mode shows comparable results in the measurement of peripapillary RNFL and choroidal thickness.
使用海德堡Spectralis光学相干断层扫描(OCT),评估增强深度成像(EDI)模式下获得的视乳头周围视网膜神经纤维层(RNFL)和脉络膜厚度,并与非EDI模式下获得的结果进行比较。
本研究纳入了25名正常健康受试者的50只眼以及20名患有不同眼部疾病的患者的32只眼。所有受试者均使用传统和EDI OCT协议,以视盘为中心进行直径3.4 mm的视乳头周围环形OCT扫描。使用有序评分量表评估RNFL和脉络膜巩膜交界处的可视化情况。采用配对t检验、组内相关系数(ICC)、95%一致性界限(LoA)以及Bland-Altman图来检验测量结果的一致性。
使用和不使用EDI获得的RNFL可见度评分无显著差异(P = 0.532),使用EDI协议时脉络膜巩膜交界处的可视化效果优于传统协议(P < 0.001)。使用EDI获得的视乳头周围RNFL厚度略厚于不使用EDI时的厚度(103.25±9.42 µm对101.87±8.78 µm,P = 0.010)。两种协议的ICC极佳,值为0.867至0.924,全球RNFL厚度的95% LoA在-10.0至7.4 µm之间。使用EDI获得的视乳头周围脉络膜厚度略薄于不使用EDI时的厚度(147.23±51.04 µm对150.90±51.84 µm,P < 0.001)。ICC同样极佳,值为0.960至0.987,全球脉络膜厚度的95% LoA在-12.5至19.8 µm之间。