Suppr超能文献

Identifying and Managing a Malpositioned Endotracheal Tube Bite Block in an Orotracheally Intubated Patient: A Case Report.

作者信息

Chou Hui-Ling, Ruan Sheng-Yuan, Wu Huey-Dong

机构信息

From the Department of Integrated Diagnostic and Therapeutics (HLC and HDW); and Department of Internal Medicine, National Taiwan University Hospital, Taipei, Taiwan (SYR and HDW).

出版信息

Medicine (Baltimore). 2016 Mar;95(12):e3161. doi: 10.1097/MD.0000000000003161.

Abstract

The universal bite block is increasingly used in orotracheally intubated patients. Here, we report a case of pilot tube dysfunction caused by a malpositioned universal bite block in an orotracheally intubated patient. We summarize the key points on identifying and managing a malpositioned universal bite block from this case and literature review.A 74-year-old woman was emergently intubated during an episode of hyperkalemia-related cardiac arrest. A universal bite block was used for fixing the endotracheal tube. After her condition stabilized, ventilator weaning was attempted; however, a positive cuff-leak test result was observed.The cuff-leak test revealed a lack of elasticity of the pilot balloon, which was completely deflated after 2 mL of air was removed. Pilot tube dysfunction was highly suspected. The bite block was slightly pulled out, and 8 mL of air was aspirated from the pilot tube. The patient was successfully extubated without stridor and respiratory distress.Our case highlighted that a malpositioned bite block may obstruct the pilot tube, causing unfavorable consequences. While fixing the bite block on an endotracheal tube, it is crucial to ensure that the takeoff point of the pilot tube is located within the C-notch of the bite block.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f322/4998396/73269bb45c34/medi-95-e3161-g001.jpg

相似文献

9
Large Iatrogenic Tracheal Injury From Attempted Endotracheal Cuff Leak Remedy: A Case Report.
A A Pract. 2019 Sep 15;13(6):225-227. doi: 10.1213/XAA.0000000000001040.

本文引用的文献

1
Pilot balloon malfunction caused by endotracheal tube bite blocker.
Respir Care. 2014 Feb;59(2):e22-4. doi: 10.4187/respcare.02474. Epub 2013 Jun 13.
7
The Universal Bite-Block: a word of caution.
Anesth Analg. 2006 Aug;103(2):495-6. doi: 10.1213/01.ANE.0000227069.01126.50.