Zhang Joni H, Macfarlane Duncan J, Sobko Tanja
Institute of Human Performance, The University of Hong Kong, Hong Kong.
Institute of Human Performance, The University of Hong Kong, Hong Kong.
J Sci Med Sport. 2016 Dec;19(12):1015-1019. doi: 10.1016/j.jsams.2016.03.004. Epub 2016 Mar 17.
A "proof-of-concept" study to examine the feasibility of wearing an Actigraph GT3X+ at the Chest (resembling a neck pendant) for physical activity measurement.
A convenience sample of 45 healthy adults (23 male, mean age 20.0±4.5 years) was included in data analysis.
Participants simultaneously wore three GT3X+ accelerometers, on the Waist, Wrist, and Chest and completed 8 bouts of slow (.67, .89, 1.11m/s), average (1.33, 1.56, 1.78m/s) and fast (2.00, 2.22m/s) walking on a treadmill. Paired t-test, correlations and absolute percentage errors (APE) of accelerometer output (vector magnitude, VM) were computed for the key pairs: Waist-Wrist; and Waist-Chest.
The Wrist-site overestimated VM to a greater extent at all speeds in comparison to the Chest. Pearson's r correlations were weaker for Waist-Wrist (<.80) in comparison to the Waist-Chest (>.85). The APE's were much lower (i.e. higher agreement) for the Chest (9.23-15.5%) compared to the Wrist (19.7-54.9%). Participants also felt the Chest-site was more acceptable than the Waist-site.
PA measurements recorded by a Chest worn GT3X+ more closely resembled PA measurements recorded at the traditional Waist site than when compared to the Wrist site. When combined with high Chest site preference, the findings of our study suggest that the Chest is a feasible site for accelerometer wear.
一项“概念验证”研究,以检验佩戴在胸部(类似颈部挂件)的Actigraph GT3X+用于身体活动测量的可行性。
对45名健康成年人(23名男性,平均年龄20.0±4.5岁)的便利样本进行数据分析。
参与者同时在腰部、手腕和胸部佩戴三个GT3X+加速度计,并在跑步机上完成8组慢走(0.67、0.89、1.11米/秒)、平均速度走(1.33、1.56、1.78米/秒)和快走(2.00、2.22米/秒)。计算关键配对(腰部-手腕;腰部-胸部)的加速度计输出(矢量大小,VM)的配对t检验、相关性和绝对百分比误差(APE)。
与胸部相比,手腕部位在所有速度下对VM的高估程度更大。与腰部-胸部(>.85)相比,腰部-手腕的Pearson's r相关性较弱(<.80)。与手腕(19.7-54.9%)相比,胸部的APE要低得多(即一致性更高)(9.23-15.5%)。参与者还感觉胸部佩戴部位比腰部佩戴部位更易接受。
与手腕部位相比,佩戴在胸部的GT3X+记录的身体活动测量值与传统腰部部位记录的身体活动测量值更相似。结合对胸部佩戴部位的高度偏好,我们的研究结果表明胸部是佩戴加速度计的可行部位。