Wilkins C J, Moores M, Hanning C D
University Department of Anaesthesia, General Hospital, Leicester.
Br J Anaesth. 1989 Apr;62(4):439-44. doi: 10.1093/bja/62.4.439.
The effects of cold-induced vasoconstriction and venous occlusion on the detection of induced hypoxaemia by four pulse oximeters were examined in 10 volunteers. In three further subjects vasoconstriction was maintained until at least one instrument failed to detect the induced hypoxaemia. Time taken to detect hypoxaemia was increased for all instruments to between two and three times the instrument's own control value for both vasoconstriction and venous engorgement (P less than 0.01). There was highly significant variation in detected minimum saturation between the instruments (P less than 0.001). One instrument failed to detect the full extent of desaturation under the experimental conditions and was more likely to fail completely to detect desaturation than the other test instruments when influenced by vasoconstriction (P less than 0.05). Significant impairment in the performance of all the instruments tested occurred in the presence of normal pulse signals. The duration of detected reductions in oxygen saturation was not significantly affected.
在10名志愿者中,研究了冷诱导血管收缩和静脉闭塞对四种脉搏血氧仪检测诱导性低氧血症的影响。在另外三名受试者中,血管收缩持续存在,直到至少有一台仪器无法检测到诱导性低氧血症。对于所有仪器,在血管收缩和静脉充血情况下,检测到低氧血症的时间增加到仪器自身对照值的两到三倍(P<0.01)。各仪器检测到的最低饱和度存在高度显著差异(P<0.001)。在实验条件下,一台仪器未能检测到完全的去饱和程度,并且在受到血管收缩影响时,比其他测试仪器更有可能完全无法检测到去饱和(P<0.05)。在存在正常脉搏信号的情况下,所有测试仪器的性能均出现显著受损。检测到的氧饱和度降低持续时间未受到显著影响。