Suppr超能文献

基于证据的使用大面积探测器扫描仪的运动学 4D-CT 研究推荐:具有尸体相关性的体模研究。

Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation.

机构信息

Service D'imagerie Guilloz, CHRU-Nancy Hôpital Central, Nancy, F-54000, France.

IADI U947, Université de Lorraine, Nancy, F-54000, France.

出版信息

Eur Radiol. 2017 Feb;27(2):437-446. doi: 10.1007/s00330-016-4362-y. Epub 2016 Apr 19.

Abstract

PROPOSE

To establish evidence-based recommendations for musculoskeletal kinematic 4D-CT on wide area-detector CT.

MATERIALS AND METHODS

In order to assess factors influencing image quality in kinematic CT studies, a phantom consisting of a polymethylmethacrylate rotating disk with round wells of different sizes was imaged with various acquisition protocols. Cadaveric acquisitions were performed on the ankle joint during motion in two different axes and at different speeds to allow validation of phantom data. Images were acquired with a 320 detector-row CT scanner and were evaluated by two readers.

RESULTS

Motion artefacts were significantly correlated with various parameters (movement axis, distance to centre, rotation speed and volume acquisition speed) (p < 0.0001). The relation between motion artefacts and distance to motion fulcrum was exponential (R 0.99). Half reconstruction led to a 23 % increase in image noise and a 40 % decrease in motion artefacts. Cadaveric acquisitions confirmed phantom data. Based on these findings, high tube rotation speed and half reconstruction are recommended for kinematic CT. The axis of motion significantly influences image artefacts and should be considered in patient training and evaluation of acquisition protocol suitability.

CONCLUSION

This study provides evidence-based recommendations for musculoskeletal kinematic 4D-CT.

KEY POINTS

• Motion artefacts can hamper the quality and interpretation of dynamic joint studies • The recommendations presented here help increase image quality • Patient training and preparation can be improved • The artefact-free distance concept helps protocol adaptation and comparison.

摘要

建议

建立基于证据的广泛探测器 CT 运动学 4D-CT 成像推荐意见。

材料和方法

为了评估运动学 CT 研究中影响图像质量的因素,我们使用一个由不同大小的圆形孔的聚甲基丙烯酸甲酯旋转盘组成的体模,采用各种采集方案进行成像。在两个不同的轴上以不同的速度对踝关节进行尸体采集,以允许验证体模数据。使用 320 排探测器 CT 扫描仪进行图像采集,并由两位读者进行评估。

结果

运动伪影与各种参数(运动轴、距中心距离、旋转速度和容积采集速度)显著相关(p<0.0001)。运动伪影与运动枢轴的距离呈指数关系(R=0.99)。半重建导致图像噪声增加 23%,运动伪影减少 40%。尸体采集证实了体模数据。基于这些发现,建议在运动学 CT 中使用高管旋转速度和半重建。运动轴对图像伪影有显著影响,应在患者培训和采集方案适用性评估中考虑。

结论

本研究为肌肉骨骼运动学 4D-CT 提供了基于证据的推荐意见。

关键点

  1. 运动伪影会影响动态关节研究的质量和解读;

  2. 本研究提出的建议有助于提高图像质量;

  3. 患者的培训和准备可以得到改善;

  4. 无伪影距离的概念有助于协议的适应和比较。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验