Department of Radiology, Yonsei University College of Medicine, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-752, Korea.
Yonsei Med J. 2013 Jul;54(4):1026-32. doi: 10.3349/ymj.2013.54.4.1026.
To validate the feasibility of real time kinematography with four-dimensional (4D) dynamic functional wrist joint imaging using dual source CT.
Two healthy volunteers performed radioulnar deviation and pronation- supination wrist motions for 10 s and 4 s per cycle in a dual source CT scanner. Scan and reconstruction protocols were set to optimize temporal resolution. Cine images of the reconstructed carpal bone of the moving wrist were recorded. The quality of the images and radiation dosage were evaluated.
The 4D cine images obtained during 4 s and 10 s of radioulnar motion showed a smooth stream of movement with good quality and little noise or artifact. Images from the pronation-supination motion showed noise with a masked surface contour. The temporal resolution was optimized at 0.28 s.
Using dual source CT, 4D cine images of in vivo kinematics of wrist joint movement were obtained and found to have a shorter scan time, improved temporal resolution and lower radiation dosages compared with those previously reported.