Ciarmatori Alberto, Nocetti L, Mistretta G, Zambelli G, Costi T
Medical Physics Department, Az. Ospedaliero - Universitaria di Modena- Policlinico, Via del Pozzo, 71, 41124, Modena, Italy.
Post Graduate School in Medical Physics, Alma Mater Studiorum University, Bologna, Italy.
Australas Phys Eng Sci Med. 2016 Jun;39(2):583-9. doi: 10.1007/s13246-016-0445-y. Epub 2016 Apr 20.
The eye lens is considered to be among the most radiosensitive human tissues. Brain CT scans may unnecessarily expose it to radiation even if the area of clinical interest is far from the eyes. The aim of this study is to implement a bismuth eye lens shielding system for Head-CT acquisitions in these cases. The study is focused on the assessment of the dosimetric characteristics of the shielding system as well as on its effect on image quality. The shielding system was tested in two set-ups which differ for distance ("contact" and "4 cm" Set up respectively). Scans were performed on a CTDI phantom and an anthropomorphic phantom. A reference set up without shielding system was acquired to establish a baseline. Image quality was assessed by signal (not HU converted), noise and contrast-to-noise ratio (CNR) evaluation. The overall dose reduction was evaluated by measuring the CTDIvol while the eye lens dose reduction was assessed by placing thermoluminescent dosimeters (TLDs) on an anthropomorphic phantom. The image quality analysis exhibits the presence of an artefact that mildly increases the CT number up to 3 cm below the shielding system. Below the artefact, the difference of the Signal and the CNR are negligible between the three different set-ups. Regarding the CTDI, the analysis demonstrates a decrease by almost 12 % (in the "contact" set-up) and 9 % (in the "4 cm" set-up). TLD measurements exhibit an eye lens dose reduction by 28.5 ± 5 and 21.1 ± 5 % respectively at the "contact" and the "4 cm" distance. No relevant artefact was found and image quality was not affected by the shielding system. Significant dose reductions were measured. These features make the shielding set-up useful for clinical implementation in both studied positions.
眼晶状体被认为是人类最具放射敏感性的组织之一。即使临床关注区域远离眼睛,脑部CT扫描也可能会使其不必要地暴露于辐射中。本研究的目的是针对这些情况,为头部CT采集实施一种铋眼晶状体屏蔽系统。该研究聚焦于评估屏蔽系统的剂量学特性及其对图像质量的影响。屏蔽系统在两种不同距离设置(分别为“接触式”和“4厘米”设置)下进行了测试。在CTDI模体和人体模体上进行扫描。采集了无屏蔽系统的参考设置以建立基线。通过信号(未转换为HU)、噪声和对比噪声比(CNR)评估来评估图像质量。通过测量CTDIvol评估总体剂量降低情况,而通过在人体模体上放置热释光剂量计(TLD)来评估眼晶状体剂量降低情况。图像质量分析显示存在一种伪影,该伪影在屏蔽系统下方3厘米处会使CT值轻度增加。在伪影下方,三种不同设置之间的信号和CNR差异可忽略不计。关于CTDI,分析表明在“接触式”设置中降低了近12%,在“4厘米”设置中降低了9%。TLD测量显示在“接触式”和“4厘米”距离处,眼晶状体剂量分别降低了28.5±5%和21.1±5%。未发现相关伪影,且屏蔽系统未影响图像质量。测量到显著的剂量降低。这些特性使屏蔽设置在两个研究位置的临床应用中都很有用。