J Invasive Cardiol. 2016 Aug;28(8):306-10. Epub 2016 Apr 15.
We prospectively examined the impact of side-holes and guide-catheter disengagement on fractional flow reserve (FFR) measurements.
Twenty-five patients undergoing clinically indicated FFR measurement for intermediate coronary artery stenosis were enrolled. Four FFR measurements were made in random order during intravenous adenosine infusion with: (a) an engaged side-hole guide catheter; (b) a disengaged side-hole guide catheter; (c) an engaged non-side-hole guide catheter; and (d) disengaged non-side-hole guide catheter.
Mean patient age was 65 ± 9 years and 100% were men. The mean distal poststenotic pressure/proximal aortic pressure (Pd/Pa) at baseline was 0.93 ± 0.05 mm Hg. Using intravenous adenosine infusion, the mean FFR measured with engaged vs disengaged non-side-hole guide catheters was 0.87 ± 0.09 vs 0.83 ± 0.10, respectively (mean difference, 0.039 ± 0.04; P<.001). The mean FFR with engaged vs disengaged side-hole guide catheters was 0.85 ± 0.10 vs 0.83 ± 0.10 (mean difference, 0.020 ± 0.02; P<.001). The mean difference in FFR measurements was 0.024 ± 0.03 (P<.001) among engaged guide catheters and 0.005 ± 0.03 (P=.47) among disengaged guide catheters.
When FFR measurements are performed with engaged guide catheters, side-hole catheters provide lower measurements. When FFR measurements are obtained with disengaged guide catheters, they are even lower and similar between guide catheter types.
我们前瞻性地研究了侧孔和导引导管脱离对血流储备分数(FFR)测量的影响。
纳入25例因中度冠状动脉狭窄而接受临床指征性FFR测量的患者。在静脉注射腺苷期间,随机顺序进行四次FFR测量,分别使用:(a)连接侧孔的导引导管;(b)脱离侧孔的导引导管;(c)连接非侧孔的导引导管;以及(d)脱离非侧孔的导引导管。
患者平均年龄为65±9岁,均为男性。基线时平均狭窄远端压力/近端主动脉压力(Pd/Pa)为0.93±0.05 mmHg。使用静脉注射腺苷时,连接与脱离非侧孔导引导管测得的平均FFR分别为0.87±0.09和0.83±0.10(平均差值为0.039±0.04;P<0.001)。连接与脱离侧孔导引导管测得的平均FFR分别为0.85±0.10和0.83±0.10(平均差值为0.020±0.02;P<0.001)。连接的导引导管之间FFR测量的平均差值为0.024±0.03(P<0.001),脱离的导引导管之间为0.005±0.03(P=0.47)。
当使用连接的导引导管进行FFR测量时,侧孔导管测得的值较低。当使用脱离的导引导管进行FFR测量时,测得值更低,且不同类型导引导管之间相似。