Suppr超能文献

超高场下多发性硬化病变成像中相位成像与定量磁化率映射的比较

A comparison of phase imaging and quantitative susceptibility mapping in the imaging of multiple sclerosis lesions at ultrahigh field.

作者信息

Cronin Matthew John, Wharton Samuel, Al-Radaideh Ali, Constantinescu Cris, Evangelou Nikos, Bowtell Richard, Gowland Penny Anne

机构信息

Brain Imaging and Analysis Centre, Duke University, Durham, NC, 27710, USA.

Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, University Park, Nottingham, NG7 2RD, UK.

出版信息

MAGMA. 2016 Jun;29(3):543-57. doi: 10.1007/s10334-016-0560-5. Epub 2016 Apr 25.

Abstract

OBJECTIVE

The aim of this study was to compare the use of high-resolution phase and QSM images acquired at ultra-high field in the investigation of multiple sclerosis (MS) lesions with peripheral rings, and to discuss their usefulness for drawing inferences about underlying tissue composition.

MATERIALS AND METHODS

Thirty-nine Subjects were scanned at 7 T, using 3D T 2*-weighted and T 1-weighted sequences. Phase images were then unwrapped and filtered, and quantitative susceptibility maps were generated using a thresholded k-space division method. Lesions were compared visually and using a 1D profiling algorithm.

RESULTS

Lesions displaying peripheral rings in the phase images were identified in 10 of the 39 subjects. Dipolar projections were apparent in the phase images outside of the extent of several of these lesions; however, QSM images showed peripheral rings without such projections. These projections appeared ring-like in a small number of phase images where no ring was observed in QSM. 1D profiles of six well-isolated example lesions showed that QSM contrast corresponds more closely to the magnitude images than phase contrast.

CONCLUSIONS

Phase images contain dipolar projections, which confounds their use in the investigation of tissue composition in MS lesions. Quantitative susceptibility maps correct these projections, providing insight into the composition of MS lesions showing peripheral rings.

摘要

目的

本研究旨在比较超高场下采集的高分辨率相位图像和定量磁敏感图(QSM)在多发性硬化(MS)外周环状病变研究中的应用,并探讨它们对于推断潜在组织成分的有用性。

材料与方法

39名受试者接受了7T扫描,采用三维T2*加权和T1加权序列。然后对相位图像进行展开和滤波,并使用阈值化k空间分割方法生成定量磁敏感图。通过视觉检查和一维轮廓分析算法对病变进行比较。

结果

39名受试者中有10名在相位图像中显示出外周环状病变。在其中一些病变范围之外的相位图像中可见偶极投影;然而,QSM图像显示外周环状病变但无此类投影。在少数QSM中未观察到环状的相位图像中,这些投影呈环状。六个隔离良好的典型病变的一维轮廓显示,与相位对比度相比,QSM对比度与幅度图像更接近。

结论

相位图像包含偶极投影,这使其在MS病变组织成分研究中的应用受到干扰。定量磁敏感图校正了这些投影,为显示外周环状的MS病变成分提供了见解。