Krag M H, Byrt W, Pope M
Department of Orthopaedics and Rehabilitation, University of Vermont College of Medicine, McClure Musculoskeletal Research Center, Burlington.
Spine (Phila Pa 1976). 1989 Mar;14(3):247-50. doi: 10.1097/00007632-198903000-00001.
Failure of attachment ("pull-off") of Gardner-Wells tongs from the cranium occasionally occurs, and may cause problems, especially in cases of significant cervical spinal instability. The optimal method for Gardner-Wells tongs attachment is not well defined: the effect on pull-off strength of stem protrusion of the spring-loaded pin, and user accuracy in setting stem protrusion do not appear to be documented in the medical literature. The authors measured pull-off strength from five fresh cadaveric crania, for each of five stem protrusion settings. In a related experiment, staff surgeons were asked to follow the standard instructions for applying clinically available Gardner-Wells tongs (smooth, unmarked indicator stems on the spring-loaded pins), after which the actual stem protrusion lengths were recorded. When the indicator stem protruded 1.0 mm (manufacturer's recommendation), the mean pull-off strength (+/- standard deviation [SD] was 137 +/- 34 pounds (610 +/- 151 N). Even with stem protrusions as low as 0.25 mm, the lowest pull-off strength was 60 pounds. Failures of fixation occurring with traction loads of 35 to 50 pounds are almost surely associated with stem protrusions of less than 0.25 mm. Even without an indicator line on the stem at 1 mm, all of the 13 orthopaedists tested produced an actual stem protrusion of 0.37 mm or more. Secure attachment of larger cranial traction loads requires careful attention to pin tightening. Proper location on the skull and the risk of penetration through the inner table must also be kept in mind.
Gardner-Wells颅骨钳有时会从颅骨上脱落(“拔出”),这可能会引发问题,尤其是在颈椎严重不稳定的情况下。Gardner-Wells颅骨钳的最佳固定方法尚未明确界定:弹簧销杆突出对拔出力的影响以及使用者设定杆突出的准确性在医学文献中似乎并无记载。作者针对五种杆突出设置,测量了来自五个新鲜尸体颅骨的拔出力。在一项相关实验中,要求外科医生按照临床可用Gardner-Wells颅骨钳(弹簧销上光滑、无标记的指示杆)的标准说明进行操作,之后记录实际的杆突出长度。当指示杆突出1.0毫米(制造商建议)时,平均拔出力(±标准差[SD])为137±34磅(610±151牛)。即使杆突出低至0.25毫米,最低拔出力也为60磅。在35至50磅的牵引负荷下发生的固定失败几乎肯定与杆突出小于0.25毫米有关。即使杆上没有1毫米的指示线,接受测试的13位骨科医生所产生的实际杆突出也均为0.37毫米或更长。要牢固固定较大的颅骨牵引负荷,需要特别注意拧紧销钉。还必须牢记在颅骨上的正确位置以及穿透内板的风险。