J Cataract Refract Surg. 2016 Apr;42(4):613-7. doi: 10.1016/j.jcrs.2015.11.048.

Abstract

PURPOSE

To compare the relative efficiency and chatter of the 30-degree 0.9 mm Kelman tip (standard tip) and Ozil Intrepid Balanced tip (torsional tip) with identical optimum settings.

SETTING

John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA.

DESIGN

Experimental study.

METHODS

Porcine nuclei were fixed in formalin and cut into 2.0 mm cubes. Lens cubes were phacoemulsified with standard and torsional tips at 60% and 100% torsional amplitude with 0%, 25%, 50%, 75%, and 100% longitudinal power with monitored forced infusion. Experiments were performed at a constant vacuum, aspiration rate, and intraocular pressure.

RESULTS

Linear regression showed increased efficiency with up to 75% longitudinal power at 60% torsional amplitude for both tips (P = .046 standard; P = .0093 torsional). Longitudinal did not significantly increase efficiency at 100% torsional power for either tip. Torsional tips were on average 29% more efficient than standard tips throughout the range tested (P < .0001), and 100% torsional power was on average 45% more efficient than 60% torsional power (P = .0028). Chatter was minimal except at 100% longitudinal power, which was greater than at 75% longitudinal power (P < .0001). At 100% longitudinal power, torsional tips had more chatter than standard tips (at 60% torsional, P < .0001; at 100% torsional, P = .0022).

CONCLUSIONS

Increasing up to 75% longitudinal power increased the efficiency at 60% torsional for standard and torsional tips. Torsional tips were more efficient than standard tips. Chatter was generally minimal. Torsional tips showed more chatter than standard tips at 100% longitudinal power.

FINANCIAL DISCLOSURE

None of the authors has a financial or proprietary interest in any material or method mentioned.

摘要

目的

比较 30 度 0.9 毫米 Kelman 尖端(标准尖端)和 Ozil Intrepid 平衡尖端(扭转尖端)在相同最佳设置下的相对效率和颤动。

设置

美国犹他州盐湖城约翰 A. 莫兰眼中心实验室。

设计

实验研究。

方法

猪核在福马林中固定,切成 2.0 毫米的立方体。晶状体方块用标准和扭转尖端在 60%和 100%扭转幅度下以 0%、25%、50%、75%和 100%的纵向功率进行超声乳化,同时进行监测强制灌输。实验在恒定的真空度、抽吸率和眼内压下进行。

结果

线性回归显示,在 60%的扭转幅度下,两种尖端的纵向功率最高可达 75%,效率都有所提高(标准:P=0.046;扭转:P=0.0093)。对于两种尖端,100%的扭转功率都没有显著提高效率。在整个测试范围内,扭转尖端的效率平均比标准尖端高 29%(P<0.0001),100%的扭转功率比 60%的扭转功率平均高 45%(P=0.0028)。除了在 100%的纵向功率下,颤动最小,而且比在 75%的纵向功率下更大(P<0.0001)。在 100%的纵向功率下,扭转尖端的颤动比标准尖端更大(在 60%的扭转时,P<0.0001;在 100%的扭转时,P=0.0022)。

结论

在 60%的扭转时,最高增加 75%的纵向功率可以提高标准和扭转尖端的效率。扭转尖端比标准尖端更有效率。颤动通常很小。在 100%的纵向功率下,扭转尖端的颤动比标准尖端更多。