高真空和抽吸对 Centurion 超声乳化效率和振动的影响。

High vacuum and aspiration on phacoemulsification efficiency and chatter for Centurion.

机构信息

Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.

Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah..

出版信息

Can J Ophthalmol. 2019 Feb;54(1):136-138. doi: 10.1016/j.jcjo.2018.03.009. Epub 2018 May 18.

Abstract

OBJECTIVE

To compare relative efficiency and chatter of high aspiration and vacuum settings.

DESIGN

In vitro laboratory study.

METHODS

The John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, was the study setting. Porcine nuclei were fixed in formalin and cut into 2.0 mm cubes. Lens cubes were phacoemulsified with Balanced tips at 50 and 60 mL/min aspiration with 500, 600, and 700 mm Hg vacuum with monitored forced infusion. Experiments were conducted at constant torsional power, longitudinal power, and intraocular pressure.

RESULTS

No significant change was observed in average chatter across each tested setting. Increasing aspiration rate did not increase efficiency. Increasing vacuum up to 600 mm Hg from 500 mm Hg did not change efficiency. However, increasing vacuum to 700 mm Hg decreased efficiency (p = 0.008 for 500 mm Hg vs 700 mm Hg and p = 0.05 for 600 mm Hg vs 700 mm Hg). Increasing aspiration and increasing vacuum did not significantly improve chatter.

CONCLUSIONS

Increasing aspiration above 50 mL/min did not improve phacoemulsification efficiency. Increasing vacuum settings to 700 mm Hg decreases efficiency. Chatter did not significantly change with increasing aspiration and vacuum settings.

摘要

目的

比较高抽吸和真空设置的相对效率和振动。

设计

体外实验室研究。

方法

犹他州盐湖城的约翰 A. 莫兰眼中心实验室是研究地点。猪核在福尔马林中固定并切成 2.0 毫米的方块。晶状体方块用平衡尖端在 50 和 60 mL/min 的抽吸下以 500、600 和 700 mmHg 的真空进行超声乳化,并监测强制输注。实验在恒定的扭转力、纵向力和眼内压下进行。

结果

在每个测试设置中,平均振动都没有明显变化。增加抽吸率不会提高效率。将真空从 500mmHg 增加到 600mmHg 不会改变效率。但是,将真空增加到 700mmHg 会降低效率(500mmHg 与 700mmHg 相比,p=0.008;600mmHg 与 700mmHg 相比,p=0.05)。增加抽吸和增加真空并没有显著改善振动。

结论

抽吸率超过 50mL/min 不会提高超声乳化效率。将真空设置增加到 700mmHg 会降低效率。振动随着抽吸和真空设置的增加没有明显变化。