Ucar Fatma, Erden Gonul, Ozdemir Seyda, Yildiz Zeynep, Arzuhal Abdullah Ercan, Temel Ismail
Clin Lab. 2016;62(3):443-9. doi: 10.7754/clin.lab.2015.150742.
The aim of our study was to assess the analytical performance of the Tosoh HLC-723G8 automated analyzer and to compare it with the Trinity Biotech Premier Hb9210 analyzer for the measurement of hemoglobin A₁c (HbA₁c).
A total of 101 patients with pre-diabetes or diabetes mellitus were included in the study. HbA₁c, was measured by both an ion-exchange high-performance liquid chromatography (IE-HPLC) method and a boronate affinity chromatography method. Statistical analysis was performed using Deming regression. Bland-Altman plots were used to calculate mean difference (bias).
The CV% values of IE-HPLC and boronate affinity methods for within run and between days were lower than 2.0%. High correlation was found (y = 1.0045x + 0.2111; r = 0.9941) between the two methods. The method shows no interference from carbamylated hemoglobin.
Both systems showed acceptable performance and are suitable for clinical application in the analysis of HbA₁c. However, laboratories should be aware of the limitations of their methods and the availability of more accurate and precise HbA₁c, determination methods.