Breen Edmond J, Tan Woei, Khan Alamgir
Australian Proteome Analysis Facility (APAF), Level 4, Building F7B, Research Park Drive, Macquarie University, Sydney NSW 2109 Australia.
Bio-Rad Laboratories Inc. 2000 Alfred Nobel Drive, Hercules, CA 94547 USA.
Sci Rep. 2016 May 31;6:26996. doi: 10.1038/srep26996.
Tissue samples (plasma, saliva, serum or urine) from 169 patients classified as either normal or having one of seven possible diseases are analysed across three 96-well plates for the presences of 37 analytes using cytokine inflammation multiplexed immunoassay panels. Censoring for concentration data caused problems for analysis of the low abundant analytes. Using fluorescence analysis over concentration based analysis allowed analysis of these low abundant analytes. Mixed-effects analysis on the resulting fluorescence and concentration responses reveals a combination of censoring and mapping the fluorescence responses to concentration values, through a 5PL curve, changed observed analyte concentrations. Simulation verifies this, by showing a dependence on the mean florescence response and its distribution on the observed analyte concentration levels. Differences from normality, in the fluorescence responses, can lead to differences in concentration estimates and unreliable probabilities for treatment effects. It is seen that when fluorescence responses are normally distributed, probabilities of treatment effects for fluorescence based t-tests has greater statistical power than the same probabilities from concentration based t-tests. We add evidence that the fluorescence response, unlike concentration values, doesn't require censoring and we show with respect to differential analysis on the fluorescence responses that background correction is not required.
使用细胞因子炎症多重免疫分析板,在三个96孔板上对169名被分类为正常或患有七种可能疾病之一的患者的组织样本(血浆、唾液、血清或尿液)进行分析,以检测37种分析物的存在情况。浓度数据的删失给低丰度分析物的分析带来了问题。使用基于荧光的分析而非基于浓度的分析能够对这些低丰度分析物进行分析。对所得荧光和浓度响应进行的混合效应分析表明,通过五参数逻辑斯谛(5PL)曲线对荧光响应进行删失和映射到浓度值的组合,改变了观察到的分析物浓度。模拟通过显示对平均荧光响应及其在观察到的分析物浓度水平上的分布的依赖性来验证这一点。荧光响应与正态性的差异可能导致浓度估计的差异以及治疗效果概率的不可靠。可以看出,当荧光响应呈正态分布时,基于荧光的t检验的治疗效果概率比基于浓度的t检验的相同概率具有更大的统计功效。我们补充证据表明,与浓度值不同,荧光响应不需要删失,并且我们通过对荧光响应的差异分析表明不需要背景校正。