Intubating introducers are often used to facilitate difficult endotracheal intubation. We report a case in which a Frova catheter was used for insertion of a laser-resistant endotracheal tube and in which a strip of plastic was shaved off from the catheter. The plastic strip formed a convolute mass, causing relevant airway obstruction. Clinicians should be aware of this potentially dangerous complication and should be cautious in using Frova intubating introducers in combination with laser-resistant tubes.
气管插管引导器常用于辅助困难气管插管。我们报告一例使用弗罗瓦导管插入抗激光气管导管的病例,在此过程中导管上一条塑料条被刮掉。这条塑料条形成了一个盘旋状团块,导致了相关气道梗阻。临床医生应意识到这种潜在的危险并发症,并且在将弗罗瓦气管插管引导器与抗激光导管联合使用时应谨慎。